ZEUS-200S/150 ZO-OM-007
REV : 08 36 ZERONE
Table 204
Guidance and manufacturer
’
s declaration
–
electromagnetic immunity
–
for equipment and systems that are not life-supporting
ZEUS-200S/150 model is intended for use in the electromagnetic environment specified below. The
customer or user of the ZEUS-200S/150 model should ensure that it is used in such an environment.
Emissions test
EN/IEC
60601
test level
Compliance
level
Electromagnetic environment - guidance
Conducted RF
IEC 61000-4-6
3 Vrms
0.15
MHz-80
MHz 6V in ISM
bands
between
0.15 MHz and 80
MHz 80% AM at
1 kJHz
3 Vrms
0.15
MHz-80
MHz 6V in ISM
bands between
0.15 MHz and
80 MHz 80%
AM at 1 kJHz
Portable and mobile RF communications
equipment should be used no closer to any part
of the ZEUS-200S/150 model, including cables,
than the recommended separation distance
calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance:
80 MHz to 800 MHz
800 MHz to 2.7GHz
Where P is the maximum output power rating of
the transmitter in watts [W] according to the
transmitter manufacturer and d is the
recommended separation in meters [m]. Field
strengths from fixed RF transmitters, as
determined by an electromagnetic site survey
a
,
should be less than the compliance level in each
frequency range
b
. Interference may occur in the
vicinity of equipment marked with the following
symbol:
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz - 2.7 GHz
80 % AM at 1 kHz
3 V/m
80 MHz - 2.7 GHz
80 % AM at 1 kHz
Note: At 80 MHz and 800 MHz, the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection by structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio [cellular/cordless] telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which
the unit is used exceeds the applicable RF compliance level above, the unit should be observed to verify
normal operation. If abnormal performance is observed, additional measures may be necessary, such as
reorienting or relocating the unit.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V1]V/m.