34
IFU-001 Rev. 8
Table 4.
Separation Distances
Recommended separation distances between portable and mobile RF communications equipment and Therm-
X
Therm-X is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of Therm-X can help prevent electromagnetic interference by maintaining
a minimum distance between portable and mobile RF communications equipment (transmitters) and Therm-X
as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power of
transmitter
(W)
Separation distance according to frequency of transmitter
(m)
150 kHz to 80 MHz
d = 1.2kH
80 MHz to 800 MHz
d = 1.2√P
800 MHz to 2.5 GHz
d = 2.3√P
0.01
.12
.12
.23
.10
.38
.38
.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacture.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
18
S
ERVICE AND
C
USTOMER
S
UPPORT
When reporting problems, provide the Unique Device Identification number (e.g. S/N) of the device
(as identified on the device label) and the affected garment, in addition to the corresponding alert
information, if any.
Contact Information:
Zenith Technical Innovations
1396 St. Paul Ave.
Gurnee, IL 60031
847.672.7481