26
IFU-001 Rev. 8
Device Physical Specifications
Parameter
Value
Dimensions of Device
Dimensions (L x W x H)
10” W x 9” H x 15” L
Weight
14lbs. maximum when empty and 15 lbs. maximum with fluid.
Umbilical Hose
(with Therm-X AT)
Length
5 ft +/- 0.5 ft
Type
3-in-1 connector for multi-patient use garments
DME Umbilical Hose
(with Therm-X Home)
Length
5 ft +/- 0.5 ft
Type
3 individual connectors for single-patient use garments
Split Umbilical Hose
(purchased separately)
Length
1 ft hose connected to two discreet sections of 4 ft +/- 0.5 ft hose
Type
Two 3-in-1 connectors for multi-patient use garments
Classifications
Information about IEC 60601-1 classification
Class of protection against
electric shock
II
Protection against accidental
contact and ingress of solid
foreign bodies
Protection against penetration of
liquids
IP21
Degree of protection against:
Touch by fingers and objects with Ø
≥ 12.5
mm
Vertically falling drops shall have no harmful effect
Degree of safety in the presence
of flammable anesthetics or
oxygen:
Not suitable for use in an oxygen enriched environment or in the
presence of flammable anesthetics
Power Supply
Type:
IEC 60601-1 compliant, 2x MOPP medical grade
Line Voltage:
100 V AC
–
240 V AC
Frequency
50/60 Hz (automatic)
Coolant
Formulation
90% Distilled Water, 10% Isopropyl Alcohol
Capacity
650 ml
Standards
Structural safety
IEC 60601-1
Summary of Contents for Therm-X Series