VICTORY RF OWNER’S MANUAL - NORTH AMERICA |
9
8
| VICTORY RF OWNER’S MANUAL - NORTH AMERICA
For the latest updates and information regarding our products, visit zeiss.com/us/sportsoptics.
ATTACHING THE CARRYING STRAP AND THE PROTECTIVE CAPS
The carrying strap
(Fig. 4/15)
and the eyepiece cap
(Fig. 4/16)
are
attached as shown in the illustrations. Note: Feed the carrying strap only
once through the eyelet on the eyepiece cap. Depending on personal
preference, use the carrying strap to connect the eyepiece cap either on
both sides or on one side only. The eyepiece cap is held onto the eyecups by
a catch. Before using the binoculars, remove the eyepiece cap with the index
finger. After observation, replace the eyepiece cap in order to protect the
eyepieces. The protective lens covers
(Fig. 5/17)
are fitted on the binoculars
as shown.
OBSERVATION WITH AND WITHOUT EYEGLASSES
When observing without glasses, use the equipment with the eyecup
extended. To do this, rotate the eyecup
(Fig. 6/1)
counterclockwise until
it locks in the highest position. The eyecup can be
set
in four positions –
in the upper and lower settings, plus in two intermediate positions. This
adjustment option allows variation of the distance from the eye to the exit
pupil and thus individual adjustment for each user. When observing
with
glasses, rotate the eyecup clockwise until it stops in the most downward
position
(Fig. 6)
.
CLEANING AND REPLACING THE EYECUPS
For cleaning purposes, the eyecups can be removed. To do so, turn the
eyecups counterclockwise past the uppermost set position, and continue to
turn until the eyecup is loosened for removal.
After proper cleaning of the eyecup and lenses, ensure there is no debris
(e.g., sand, dirt, or grime) on the threads of the eyecup housing and ocular
housing. Replace the eyecups by turning them onto the ocular housing in a
clockwise direction. Be careful not to cross-thread the fine threads. Continue
to turn in a clockwise direction to the lowest setting position, and apply
a small amount of additional torque to secure the eyecup to the ocular
housing. The eyecups are ready for regular adjustments at this time.
4 | 7
EN
Adjusting the interpupilary distance
By folding the halves of the binoculars about the central axis, the eye relief can
be adjusted such that a circular image is formed when observing with both eyes
(Fig. 7)
.
Note:
Depending on the set of the interpupilary distance, the aiming mark and
display may be slightly angled.
Focusing the aiming mark and diopter compensation
Use the range finder button
(Fig. 1/6)
to switch on the aiming mark
(Fig. 2/9)
and
hold down the button. Focus the aiming mark and the display by turning the
right
hand diopter compensation
(Fig. 1/4)
to the left or the right. Then carefully use
the central focusing
(Fig. 1/2)
for sharp focusing of the image in the right hand
binocular tube.
Next, use the
left
hand diopter compensation
(Fig. 1/3)
for sharp focusing on the
same object of the image in the left hand binocular tube. The values that are set can
be read from the “+” or “–” scale on the back of the binoculars.
Menu settings in the rangefinder
Press the SET button for approx. two seconds to load the menu of your ZEISS Victory
RF. You can then switch between the individual menu options by pressing the SET
button once for each option.
Table 1
Menu 1
Brightness
Press SET button for 2 seconds
Menu 2
Ballistics settings
Press SET button for 2 seconds
+ press the SET button once
Menu 3
Display settings
Press SET button for 2 seconds
+ press the SET button twice
Menu 4
Unit of measure
Press SET button for 2 seconds
+ press the SET button three times
Menu 5
Measuring mode
Press SET button for 2 seconds
+ press the SET button four times
Menu 6
Key layout
Press SET button for 2 seconds
+ press the SET button five times
Menu 7
Turn off
Press SET button for 2 seconds
+ press the SET button six times
Brightness
The brightness settings are located in menu 1 (press the SET button for 2 seconds).
The ZEISS Victory RF has 11 brightness curves. By pressing the range finding button,
you can select one of the 11 different brightness curves (curve 1 darkest setting
possibility; curve 11 lightest setting possibility).
In addition to this, your ZEISS Victory RF also has automatic brightness adjustment
for each of the selected brightness curves.
Simply release the range finding button
(Fig. 1/6)
when it reaches the appropriate
setting for you. The brightness curve displayed most recently will then be saved. To
check, you can always display the saved settings in the menu.
Fig. 7
Fig. 8
(2)
(4)
(1)
(3)
L
R
L
R
L
R
(2)
(4)
R
R
Fig. 6
(1)
Observation with and without glasses
When observing
without
glasses, use the equipment with the eye cup extended.
To do this, turn the eye cup
(Fig. 6/1)
upwards and to the left (anticlockwise) until it
locks in the highest position.
The eye cup can be
locked
in four positions – in the upper and lower plus in two
intermediate positions. This adjustment option allows variation of the distance from
the eye to the exit pupil and thus individual adjustment for each user.
When observing
with
glasses, turn the eye cup downwards and to the right
(clockwise) until it locks in the lowest position
(Fig. 6)
.
Cleaning and replacing the eyecups
To replace the eyecups or for cleaning, they can be completely screwed off of the
binoculars. Turn the eyecups out as shown in
Fig. 6
to the top click stop position
and turn one more thread in the same direction until the eyecup is completely
screwed out.
After cleaning, or replacing the eyecup, turn it to the right (clockwise) completely on
the eyepiece. Using a slight jerk to the right the thread of the eyepiece clicks into the
lowest setting. Finally, you can again set your desired distance between the eye and
eyepiece using the click stops.
4 | 7
EN
Adjusting the interpupilary distance
By folding the halves of the binoculars about the central axis, the eye relief can
be adjusted such that a circular image is formed when observing with both eyes
(Fig. 7)
.
Note:
Depending on the set of the interpupilary distance, the aiming mark and
display may be slightly angled.
Focusing the aiming mark and diopter compensation
Use the range finder button
(Fig. 1/6)
to switch on the aiming mark
(Fig. 2/9)
and
hold down the button. Focus the aiming mark and the display by turning the
right
hand diopter compensation
(Fig. 1/4)
to the left or the right. Then carefully use
the central focusing
(Fig. 1/2)
for sharp focusing of the image in the right hand
binocular tube.
Next, use the
left
hand diopter compensation
(Fig. 1/3)
for sharp focusing on the
same object of the image in the left hand binocular tube. The values that are set can
be read from the “+” or “–” scale on the back of the binoculars.
Menu settings in the rangefinder
Press the SET button for approx. two seconds to load the menu of your ZEISS Victory
RF. You can then switch between the individual menu options by pressing the SET
button once for each option.
Table 1
Menu 1
Brightness
Press SET button for 2 seconds
Menu 2
Ballistics settings
Press SET button for 2 seconds
+ press the SET button once
Menu 3
Display settings
Press SET button for 2 seconds
+ press the SET button twice
Menu 4
Unit of measure
Press SET button for 2 seconds
+ press the SET button three times
Menu 5
Measuring mode
Press SET button for 2 seconds
+ press the SET button four times
Menu 6
Key layout
Press SET button for 2 seconds
+ press the SET button five times
Menu 7
Turn off
Press SET button for 2 seconds
+ press the SET button six times
Brightness
The brightness settings are located in menu 1 (press the SET button for 2 seconds).
The ZEISS Victory RF has 11 brightness curves. By pressing the range finding button,
you can select one of the 11 different brightness curves (curve 1 darkest setting
possibility; curve 11 lightest setting possibility).
In addition to this, your ZEISS Victory RF also has automatic brightness adjustment
for each of the selected brightness curves.
Simply release the range finding button
(Fig. 1/6)
when it reaches the appropriate
setting for you. The brightness curve displayed most recently will then be saved. To
check, you can always display the saved settings in the menu.
Fig. 7
Fig. 8
(2)
(4)
(1)
(3)
L
R
L
R
L
R
(2)
(4)
R
R
Fig. 6
(1)
Observation with and without glasses
When observing
without
glasses, use the equipment with the eye cup extended.
To do this, turn the eye cup
(Fig. 6/1)
upwards and to the left (anticlockwise) until it
locks in the highest position.
The eye cup can be
locked
in four positions – in the upper and lower plus in two
intermediate positions. This adjustment option allows variation of the distance from
the eye to the exit pupil and thus individual adjustment for each user.
When observing
with
glasses, turn the eye cup downwards and to the right
(clockwise) until it locks in the lowest position
(Fig. 6)
.
Cleaning and replacing the eyecups
To replace the eyecups or for cleaning, they can be completely screwed off of the
binoculars. Turn the eyecups out as shown in
Fig. 6
to the top click stop position
and turn one more thread in the same direction until the eyecup is completely
screwed out.
After cleaning, or replacing the eyecup, turn it to the right (clockwise) completely on
the eyepiece. Using a slight jerk to the right the thread of the eyepiece clicks into the
lowest setting. Finally, you can again set your desired distance between the eye and
eyepiece using the click stops.
4 | 7
EN
Adjusting the interpupilary distance
By folding the halves of the binoculars about the central axis, the eye relief can
be adjusted such that a circular image is formed when observing with both eyes
(Fig. 7)
.
Note:
Depending on the set of the interpupilary distance, the aiming mark and
display may be slightly angled.
Focusing the aiming mark and diopter compensation
Use the range finder button
(Fig. 1/6)
to switch on the aiming mark
(Fig. 2/9)
and
hold down the button. Focus the aiming mark and the display by turning the
right
hand diopter compensation
(Fig. 1/4)
to the left or the right. Then carefully use
the central focusing
(Fig. 1/2)
for sharp focusing of the image in the right hand
binocular tube.
Next, use the
left
hand diopter compensation
(Fig. 1/3)
for sharp focusing on the
same object of the image in the left hand binocular tube. The values that are set can
be read from the “+” or “–” scale on the back of the binoculars.
Menu settings in the rangefinder
Press the SET button for approx. two seconds to load the menu of your ZEISS Victory
RF. You can then switch between the individual menu options by pressing the SET
button once for each option.
Table 1
Menu 1
Brightness
Press SET button for 2 seconds
Menu 2
Ballistics settings
Press SET button for 2 seconds
+ press the SET button once
Menu 3
Display settings
Press SET button for 2 seconds
+ press the SET button twice
Menu 4
Unit of measure
Press SET button for 2 seconds
+ press the SET button three times
Menu 5
Measuring mode
Press SET button for 2 seconds
+ press the SET button four times
Menu 6
Key layout
Press SET button for 2 seconds
+ press the SET button five times
Menu 7
Turn off
Press SET button for 2 seconds
+ press the SET button six times
Brightness
The brightness settings are located in menu 1 (press the SET button for 2 seconds).
The ZEISS Victory RF has 11 brightness curves. By pressing the range finding button,
you can select one of the 11 different brightness curves (curve 1 darkest setting
possibility; curve 11 lightest setting possibility).
In addition to this, your ZEISS Victory RF also has automatic brightness adjustment
for each of the selected brightness curves.
Simply release the range finding button
(Fig. 1/6)
when it reaches the appropriate
setting for you. The brightness curve displayed most recently will then be saved. To
check, you can always display the saved settings in the menu.
Fig. 7
Fig. 8
(2)
(4)
(1)
(3)
L
R
L
R
L
R
(2)
(4)
R
R
Fig. 6
(1)
Observation with and without glasses
When observing
without
glasses, use the equipment with the eye cup extended.
To do this, turn the eye cup
(Fig. 6/1)
upwards and to the left (anticlockwise) until it
locks in the highest position.
The eye cup can be
locked
in four positions – in the upper and lower plus in two
intermediate positions. This adjustment option allows variation of the distance from
the eye to the exit pupil and thus individual adjustment for each user.
When observing
with
glasses, turn the eye cup downwards and to the right
(clockwise) until it locks in the lowest position
(Fig. 6)
.
Cleaning and replacing the eyecups
To replace the eyecups or for cleaning, they can be completely screwed off of the
binoculars. Turn the eyecups out as shown in
Fig. 6
to the top click stop position
and turn one more thread in the same direction until the eyecup is completely
screwed out.
After cleaning, or replacing the eyecup, turn it to the right (clockwise) completely on
the eyepiece. Using a slight jerk to the right the thread of the eyepiece clicks into the
lowest setting. Finally, you can again set your desired distance between the eye and
eyepiece using the click stops.
3 | 7
EN
PREPARATION
Inserting/Removing the battery
The laser range finder is powered by a type CR 2 lithium battery.
To insert and replace the battery, unscrew the battery cover
(Fig. 3/8)
– using a coin
or similar – by turning anticlockwise. Insert the battery with the positive end forward
(according to the symbol in the battery compartment).
Then screw the battery cover back on by turning clockwise.
At 20 °C, a new battery will last for over 2,500 measurements.
Depending on the conditions of use, however, low temperatures or frequent use of
the scan mode for example may result in the battery life being considerably shorter.
Low battery is indicated by the appearance of the battery symbol on the display.
If the equipment will be unused for a long period, remove the battery in order to
prevent damage by leakage from the battery. Use only high quality brand batteries.
+
–
(8)
Fig. 3
Fig. 4
(16)
(15)
Fig. 5
(17)
Attaching the carrying strap and the protective caps
The carrying strap
(Fig. 4/15)
and the eyepiece cap
(Fig. 4/16)
are attached as
shown in the illustrations.
Note:
Feed the carrying strap only once through the eyelet on the eyepiece cap.
Depending on personal taste, use the carrying strap to connect the eyepiece cap
either on both sides or on one side only. The eyepiece cap is held onto the eye cups
by a catch.
Before using the binoculars, remove the eyepiece cap with the index finger.
After observation, replace the eyepiece cap in order to protect the eyepieces.
The protective lens covers
(Fig. 5/17)
are fitted on the binoculars as shown.
TECHNICAL DATA
8x42
10x42
8x54
10x54
Magnification
8
10
8
10
Effective lens diameter
mm
42
42
54
54
Exit pupil diameter
mm
5.3
4.2
6.8
5.4
Twilight factor
18.3
20.5
20.8
23.2
Field of view
m/1000m (ft/1000yds)
135 (405)
115 (345)
120 (360)
110 (330)
Subjective angle of view
°
62
66
55
63
Close-up setting limit
m (ft)
2.5 (8.2)
3.5 (11.5)
Diopter adjustment range
dpt
+/- 3
+/- 3
Eye relief
mm
17 (0.7)
14 (0.6)
Pupil Distance
mm
53.5 - 76
58.5 - 76
Lens type
FL
FL
Prism system
Abbe-König
Abbe-König
Coating
T*/LotuTec
®
T*/LotuTec
®
Nitrogen filling
Waterproof
mbar
400
400
Operating temperature
1
°C (°F)
-25 / +63 (-13 / +145)
-25 / +63 (-13 / +145)
Length
mm (in)
166 (6.5)
166 (6.5)
195 (7.7)
195 (7.7)
Width with an eye spacing of 65 mm
mm (in)
121 (4.8)
121 (4.8)
136 (5.4)
136 (5.4)
Weight
g (oz)
895 (31.6)
915 (32.3)
1,095 (38.6)
1,115 (39.3)
Measuring range
2
m (yds)
15 - 2,300 (16 - 2,500)
15 - 2,300 (16 - 2,500)
Measuring precision
mrad
± 1 – 600 / ± 0.5 % > 600
± 1 – 600 / ± 0.5 % > 600
Measuring time
Sek. (sec.)
< 0.3
< 0.3
Laser shaft length
nm
905
905
Laser beam divergence
mrad
1.6 x 0.5
1.6 x 0.5
Battery
1 x 3V / CR2
1 x 3V / CR2
Battery life at +20°C
> 2,500x
> 2,500x
Subject to changes in design and scope of supply due to technical improvements.
1
Battery is subject to a lower temperature limit.
Product also functions under – 10 °C /14 °F.
2
The range is influenced by the size and the degree of reflection of the object
as well as by the weather and sun beams.
3 | 7
EN
PREPARATION
Inserting/Removing the battery
The laser range finder is powered by a type CR 2 lithium battery.
To insert and replace the battery, unscrew the battery cover
(Fig. 3/8)
– using a coin
or similar – by turning anticlockwise. Insert the battery with the positive end forward
(according to the symbol in the battery compartment).
Then screw the battery cover back on by turning clockwise.
At 20 °C, a new battery will last for over 2,500 measurements.
Depending on the conditions of use, however, low temperatures or frequent use of
the scan mode for example may result in the battery life being considerably shorter.
Low battery is indicated by the appearance of the battery symbol on the display.
If the equipment will be unused for a long period, remove the battery in order to
prevent damage by leakage from the battery. Use only high quality brand batteries.
+
–
(8)
Fig. 3
Fig. 4
(16)
(15)
Fig. 5
(17)
Attaching the carrying strap and the protective caps
The carrying strap
(Fig. 4/15)
and the eyepiece cap
(Fig. 4/16)
are attached as
shown in the illustrations.
Note:
Feed the carrying strap only once through the eyelet on the eyepiece cap.
Depending on personal taste, use the carrying strap to connect the eyepiece cap
either on both sides or on one side only. The eyepiece cap is held onto the eye cups
by a catch.
Before using the binoculars, remove the eyepiece cap with the index finger.
After observation, replace the eyepiece cap in order to protect the eyepieces.
The protective lens covers
(Fig. 5/17)
are fitted on the binoculars as shown.
TECHNICAL DATA
8x42
10x42
8x54
10x54
Magnification
8
10
8
10
Effective lens diameter
mm
42
42
54
54
Exit pupil diameter
mm
5.3
4.2
6.8
5.4
Twilight factor
18.3
20.5
20.8
23.2
Field of view
m/1000m (ft/1000yds)
135 (405)
115 (345)
120 (360)
110 (330)
Subjective angle of view
°
62
66
55
63
Close-up setting limit
m (ft)
2.5 (8.2)
3.5 (11.5)
Diopter adjustment range
dpt
+/- 3
+/- 3
Eye relief
mm
17 (0.7)
14 (0.6)
Pupil Distance
mm
53.5 - 76
58.5 - 76
Lens type
FL
FL
Prism system
Abbe-König
Abbe-König
Coating
T*/LotuTec
®
T*/LotuTec
®
Nitrogen filling
Waterproof
mbar
400
400
Operating temperature
1
°C (°F)
-25 / +63 (-13 / +145)
-25 / +63 (-13 / +145)
Length
mm (in)
166 (6.5)
166 (6.5)
195 (7.7)
195 (7.7)
Width with an eye spacing of 65 mm
mm (in)
121 (4.8)
121 (4.8)
136 (5.4)
136 (5.4)
Weight
g (oz)
895 (31.6)
915 (32.3)
1,095 (38.6)
1,115 (39.3)
Measuring range
2
m (yds)
15 - 2,300 (16 - 2,500)
15 - 2,300 (16 - 2,500)
Measuring precision
mrad
± 1 – 600 / ± 0.5 % > 600
± 1 – 600 / ± 0.5 % > 600
Measuring time
Sek. (sec.)
< 0.3
< 0.3
Laser shaft length
nm
905
905
Laser beam divergence
mrad
1.6 x 0.5
1.6 x 0.5
Battery
1 x 3V / CR2
1 x 3V / CR2
Battery life at +20°C
> 2,500x
> 2,500x
Subject to changes in design and scope of supply due to technical improvements.
1
Battery is subject to a lower temperature limit.
Product also functions under – 10 °C /14 °F.
2
The range is influenced by the size and the degree of reflection of the object
as well as by the weather and sun beams.