background image

VICTORY RF OWNER’S MANUAL - NORTH AMERICA | 

9

8

 | VICTORY RF OWNER’S MANUAL - NORTH AMERICA

For the latest updates and information regarding our products, visit zeiss.com/us/sportsoptics.

ATTACHING THE CARRYING STRAP AND THE PROTECTIVE CAPS 

The carrying strap 

(Fig. 4/15)

 and the eyepiece cap 

(Fig. 4/16)

 are 

attached as shown in the illustrations. Note: Feed the carrying strap only 
once through the eyelet on the eyepiece cap. Depending on personal 
preference, use the carrying strap to connect the eyepiece cap either on 
both sides or on one side only. The eyepiece cap is held onto the eyecups by 
a catch. Before using the binoculars, remove the eyepiece cap with the index 

finger. After observation, replace the eyepiece cap in order to protect the 

eyepieces. The protective lens covers 

(Fig. 5/17)

 are fitted on the binoculars 

as shown.

OBSERVATION WITH AND WITHOUT EYEGLASSES

When observing without glasses, use the equipment with the eyecup 
extended. To do this, rotate the eyecup 

(Fig. 6/1)

 counterclockwise until 

it locks in the highest position. The eyecup can be 

set

 in four positions – 

in the upper and lower settings, plus in two intermediate positions. This 
adjustment option allows variation of the distance from the eye to the exit 
pupil and thus individual adjustment for each user. When observing 

with

 

glasses, rotate the eyecup clockwise until it stops in the most downward 
position

 (Fig. 6)

.

CLEANING AND REPLACING THE EYECUPS 

For cleaning purposes, the eyecups can be removed. To do so, turn the 
eyecups counterclockwise past the uppermost set position, and continue to 
turn until the eyecup is loosened for removal. 

After proper cleaning of the eyecup and lenses, ensure there is no debris 
(e.g., sand, dirt, or grime) on the threads of the eyecup housing and ocular 
housing. Replace the eyecups by turning them onto the ocular housing in a 

clockwise direction. Be careful not to cross-thread the fine threads. Continue 

to turn in a clockwise direction to the lowest setting position, and apply 
a small amount of additional torque to secure the eyecup to the ocular 
housing. The eyecups are ready for regular adjustments at this time.

4 | 7 

EN

Adjusting the interpupilary distance

By folding the halves of the binoculars about the central axis, the eye relief can  
be adjusted such that a circular image is formed when observing with both eyes  

(Fig. 7)

.

Note:

 Depending on the set of the interpupilary distance, the aiming mark and 

display may be slightly angled.

Focusing the aiming mark and diopter compensation

Use the range finder button 

(Fig. 1/6)

 to switch on the aiming mark 

(Fig. 2/9)

 and 

hold down the button. Focus the aiming mark and the display by turning the 

right

 

hand diopter compensation 

(Fig. 1/4)

 to the left or the right. Then carefully use 

the central focusing 

(Fig. 1/2)

 for sharp focusing of the image in the right hand 

binocular tube. 
Next, use the 

left

 hand diopter compensation 

(Fig. 1/3)

 for sharp focusing on the 

same object of the image in the left hand binocular tube. The values that are set can 
be read from the “+” or “–” scale on the back of the binoculars.

Menu settings in the rangefinder

Press the SET button for approx. two seconds to load the menu of your ZEISS Victory 
RF. You can then switch between the individual menu options by pressing the SET 
button once for each option.

Table 1

Menu 1

Brightness

Press SET button for 2 seconds

Menu 2

Ballistics settings

Press SET button for 2 seconds  
+ press the SET button once

Menu 3

Display settings

Press SET button for 2 seconds  
+ press the SET button twice

Menu 4

Unit of measure

Press SET button for 2 seconds  
+ press the SET button three times

Menu 5

Measuring mode 

Press SET button for 2 seconds  
+ press the SET button four times

Menu 6

Key layout

Press SET button for 2 seconds  
+ press the SET button five times

Menu 7

Turn off

Press SET button for 2 seconds  
+ press the  SET button six times

Brightness

The brightness settings are located in menu 1 (press the SET button for 2 seconds). 
The ZEISS Victory RF has 11 brightness curves. By pressing the range finding button, 
you can select one of the 11 different brightness curves (curve 1 — darkest setting 
possibility; curve 11 — lightest setting possibility).
In addition to this, your ZEISS Victory RF also has automatic brightness adjustment 
for each of the selected brightness curves.
Simply release the range finding button 

(Fig. 1/6)

 when it reaches the appropriate 

setting for you. The brightness curve displayed most recently will then be saved. To 
check, you can always display the saved settings in the menu.

Fig. 7

Fig. 8

(2)

(4)

(1)

(3)

L

R

L

R

L

R

(2)

(4)

R

R

Fig. 6

(1)

Observation with and without glasses

When observing 

without

 glasses, use the equipment with the eye cup extended. 

To do this, turn the eye cup 

(Fig. 6/1)

 upwards and to the left (anticlockwise) until it 

locks in the highest position.
The eye cup can be 

locked

 in four positions – in the upper and lower plus in two 

intermediate positions. This adjustment option allows variation of the distance from
the eye to the exit pupil and thus individual adjustment for each user.
When observing 

with

 glasses, turn the eye cup downwards and to the right 

(clockwise) until it locks in the lowest position 

(Fig. 6)

.

Cleaning and replacing the eyecups

To replace the eyecups or for cleaning, they can be completely screwed off of the 
binoculars. Turn the eyecups out as shown in 

Fig. 6

 to the top click stop position 

and turn one more thread in the same direction until the eyecup is completely 
screwed out.
After cleaning, or replacing the eyecup, turn it to the right (clockwise) completely on 
the eyepiece. Using a slight jerk to the right the thread of the eyepiece clicks into the 
lowest setting. Finally, you can again set your desired distance between the eye and 
eyepiece using the click stops.

4 | 7 

EN

Adjusting the interpupilary distance

By folding the halves of the binoculars about the central axis, the eye relief can  
be adjusted such that a circular image is formed when observing with both eyes  

(Fig. 7)

.

Note:

 Depending on the set of the interpupilary distance, the aiming mark and 

display may be slightly angled.

Focusing the aiming mark and diopter compensation

Use the range finder button 

(Fig. 1/6)

 to switch on the aiming mark 

(Fig. 2/9)

 and 

hold down the button. Focus the aiming mark and the display by turning the 

right

 

hand diopter compensation 

(Fig. 1/4)

 to the left or the right. Then carefully use 

the central focusing 

(Fig. 1/2)

 for sharp focusing of the image in the right hand 

binocular tube. 
Next, use the 

left

 hand diopter compensation 

(Fig. 1/3)

 for sharp focusing on the 

same object of the image in the left hand binocular tube. The values that are set can 
be read from the “+” or “–” scale on the back of the binoculars.

Menu settings in the rangefinder

Press the SET button for approx. two seconds to load the menu of your ZEISS Victory 
RF. You can then switch between the individual menu options by pressing the SET 
button once for each option.

Table 1

Menu 1

Brightness

Press SET button for 2 seconds

Menu 2

Ballistics settings

Press SET button for 2 seconds  
+ press the SET button once

Menu 3

Display settings

Press SET button for 2 seconds  
+ press the SET button twice

Menu 4

Unit of measure

Press SET button for 2 seconds  
+ press the SET button three times

Menu 5

Measuring mode 

Press SET button for 2 seconds  
+ press the SET button four times

Menu 6

Key layout

Press SET button for 2 seconds  
+ press the SET button five times

Menu 7

Turn off

Press SET button for 2 seconds  
+ press the  SET button six times

Brightness

The brightness settings are located in menu 1 (press the SET button for 2 seconds). 
The ZEISS Victory RF has 11 brightness curves. By pressing the range finding button, 
you can select one of the 11 different brightness curves (curve 1 — darkest setting 
possibility; curve 11 — lightest setting possibility).
In addition to this, your ZEISS Victory RF also has automatic brightness adjustment 
for each of the selected brightness curves.
Simply release the range finding button 

(Fig. 1/6)

 when it reaches the appropriate 

setting for you. The brightness curve displayed most recently will then be saved. To 
check, you can always display the saved settings in the menu.

Fig. 7

Fig. 8

(2)

(4)

(1)

(3)

L

R

L

R

L

R

(2)

(4)

R

R

Fig. 6

(1)

Observation with and without glasses

When observing 

without

 glasses, use the equipment with the eye cup extended. 

To do this, turn the eye cup 

(Fig. 6/1)

 upwards and to the left (anticlockwise) until it 

locks in the highest position.
The eye cup can be 

locked

 in four positions – in the upper and lower plus in two 

intermediate positions. This adjustment option allows variation of the distance from
the eye to the exit pupil and thus individual adjustment for each user.
When observing 

with

 glasses, turn the eye cup downwards and to the right 

(clockwise) until it locks in the lowest position 

(Fig. 6)

.

Cleaning and replacing the eyecups

To replace the eyecups or for cleaning, they can be completely screwed off of the 
binoculars. Turn the eyecups out as shown in 

Fig. 6

 to the top click stop position 

and turn one more thread in the same direction until the eyecup is completely 
screwed out.
After cleaning, or replacing the eyecup, turn it to the right (clockwise) completely on 
the eyepiece. Using a slight jerk to the right the thread of the eyepiece clicks into the 
lowest setting. Finally, you can again set your desired distance between the eye and 
eyepiece using the click stops.

4 | 7 

EN

Adjusting the interpupilary distance

By folding the halves of the binoculars about the central axis, the eye relief can  
be adjusted such that a circular image is formed when observing with both eyes  

(Fig. 7)

.

Note:

 Depending on the set of the interpupilary distance, the aiming mark and 

display may be slightly angled.

Focusing the aiming mark and diopter compensation

Use the range finder button 

(Fig. 1/6)

 to switch on the aiming mark 

(Fig. 2/9)

 and 

hold down the button. Focus the aiming mark and the display by turning the 

right

 

hand diopter compensation 

(Fig. 1/4)

 to the left or the right. Then carefully use 

the central focusing 

(Fig. 1/2)

 for sharp focusing of the image in the right hand 

binocular tube. 
Next, use the 

left

 hand diopter compensation 

(Fig. 1/3)

 for sharp focusing on the 

same object of the image in the left hand binocular tube. The values that are set can 
be read from the “+” or “–” scale on the back of the binoculars.

Menu settings in the rangefinder

Press the SET button for approx. two seconds to load the menu of your ZEISS Victory 
RF. You can then switch between the individual menu options by pressing the SET 
button once for each option.

Table 1

Menu 1

Brightness

Press SET button for 2 seconds

Menu 2

Ballistics settings

Press SET button for 2 seconds  
+ press the SET button once

Menu 3

Display settings

Press SET button for 2 seconds  
+ press the SET button twice

Menu 4

Unit of measure

Press SET button for 2 seconds  
+ press the SET button three times

Menu 5

Measuring mode 

Press SET button for 2 seconds  
+ press the SET button four times

Menu 6

Key layout

Press SET button for 2 seconds  
+ press the SET button five times

Menu 7

Turn off

Press SET button for 2 seconds  
+ press the  SET button six times

Brightness

The brightness settings are located in menu 1 (press the SET button for 2 seconds). 
The ZEISS Victory RF has 11 brightness curves. By pressing the range finding button, 
you can select one of the 11 different brightness curves (curve 1 — darkest setting 
possibility; curve 11 — lightest setting possibility).
In addition to this, your ZEISS Victory RF also has automatic brightness adjustment 
for each of the selected brightness curves.
Simply release the range finding button 

(Fig. 1/6)

 when it reaches the appropriate 

setting for you. The brightness curve displayed most recently will then be saved. To 
check, you can always display the saved settings in the menu.

Fig. 7

Fig. 8

(2)

(4)

(1)

(3)

L

R

L

R

L

R

(2)

(4)

R

R

Fig. 6

(1)

Observation with and without glasses

When observing 

without

 glasses, use the equipment with the eye cup extended. 

To do this, turn the eye cup 

(Fig. 6/1)

 upwards and to the left (anticlockwise) until it 

locks in the highest position.
The eye cup can be 

locked

 in four positions – in the upper and lower plus in two 

intermediate positions. This adjustment option allows variation of the distance from
the eye to the exit pupil and thus individual adjustment for each user.
When observing 

with

 glasses, turn the eye cup downwards and to the right 

(clockwise) until it locks in the lowest position 

(Fig. 6)

.

Cleaning and replacing the eyecups

To replace the eyecups or for cleaning, they can be completely screwed off of the 
binoculars. Turn the eyecups out as shown in 

Fig. 6

 to the top click stop position 

and turn one more thread in the same direction until the eyecup is completely 
screwed out.
After cleaning, or replacing the eyecup, turn it to the right (clockwise) completely on 
the eyepiece. Using a slight jerk to the right the thread of the eyepiece clicks into the 
lowest setting. Finally, you can again set your desired distance between the eye and 
eyepiece using the click stops.

3 | 7 

EN

PREPARATION

Inserting/Removing the battery

The laser range finder is powered by a type CR 2 lithium battery.
To insert and replace the battery, unscrew the battery cover 

(Fig. 3/8)

 – using a coin 

or similar – by turning anticlockwise. Insert the battery with the positive end forward 
(according to the symbol in the battery compartment).
Then screw the battery cover back on by turning clockwise.
At 20 °C, a new battery will last for over 2,500 measurements. 
Depending on the conditions of use, however, low temperatures or frequent use of 
the scan mode for example may result in the battery life being considerably shorter.
Low battery is indicated by the appearance of the battery symbol on the display. 
If the equipment will be unused for a long period, remove the battery in order to 
prevent damage by leakage from the battery. Use only high quality brand batteries.

+

(8)

Fig. 3

Fig. 4

(16)

(15)

Fig. 5

(17)

Attaching the carrying strap and the protective caps

The carrying strap 

(Fig. 4/15)

 and the eyepiece cap 

(Fig. 4/16)

 are attached as 

shown in the illustrations.

Note:

 Feed the carrying strap only once through the eyelet on the eyepiece cap.

Depending on personal taste, use the carrying strap to connect the eyepiece cap 
either on both sides or on one side only. The eyepiece cap is held onto the eye cups
by a catch.
Before using the binoculars, remove the eyepiece cap with the index finger.
After observation, replace the eyepiece cap in order to protect the eyepieces.
The protective lens covers 

(Fig. 5/17)

 are fitted on the binoculars as shown.

TECHNICAL DATA 

8x42

10x42

8x54

10x54

Magnification

8

10

8

10

Effective lens diameter

mm

42

42

54

54

Exit pupil diameter

mm

5.3

4.2

6.8

5.4

Twilight factor

18.3

20.5

20.8

23.2

Field of view

m/1000m (ft/1000yds)

135 (405)

115 (345)

120 (360)

110 (330)

Subjective angle of view

°

62

66

55

63

Close-up setting limit

m (ft)

2.5 (8.2)

3.5 (11.5)

Diopter adjustment range

dpt

+/- 3

+/- 3

Eye relief

mm

17 (0.7)

14 (0.6)

Pupil Distance

mm

53.5 - 76

58.5 - 76

Lens type

FL

FL

Prism system

Abbe-König

Abbe-König

Coating

T*/LotuTec

®

T*/LotuTec

®

Nitrogen filling

Waterproof

mbar

400

400

Operating temperature

1

°C (°F)

-25 / +63 (-13 / +145)

-25 / +63 (-13 / +145)

Length

mm (in)

166 (6.5)

166 (6.5)

195 (7.7)

195 (7.7)

Width with an eye spacing of 65 mm

mm (in)

121 (4.8)

121 (4.8)

136 (5.4)

136 (5.4)

Weight

g (oz)

895 (31.6)

915 (32.3)

1,095 (38.6)

1,115 (39.3)

Measuring range

2

m (yds)

15 - 2,300 (16 - 2,500)

15 - 2,300 (16 - 2,500)

Measuring precision

mrad

± 1 – 600 / ± 0.5 % > 600

± 1 – 600 / ± 0.5 % > 600

Measuring time

Sek. (sec.)

< 0.3 

< 0.3 

Laser shaft length

nm

905

905

Laser beam divergence

mrad

1.6 x 0.5

1.6 x 0.5

Battery

1 x 3V / CR2

1 x 3V / CR2

Battery life at +20°C

> 2,500x

> 2,500x

Subject to changes in design and scope of supply due to technical improvements.

1

  Battery is subject to a lower temperature limit. 

Product also functions under – 10 °C /14 °F. 

2

  The range is influenced by the size and the degree of reflection of the object  

as well as by the weather and sun beams.

3 | 7 

EN

PREPARATION

Inserting/Removing the battery

The laser range finder is powered by a type CR 2 lithium battery.
To insert and replace the battery, unscrew the battery cover 

(Fig. 3/8)

 – using a coin 

or similar – by turning anticlockwise. Insert the battery with the positive end forward 
(according to the symbol in the battery compartment).
Then screw the battery cover back on by turning clockwise.
At 20 °C, a new battery will last for over 2,500 measurements. 
Depending on the conditions of use, however, low temperatures or frequent use of 
the scan mode for example may result in the battery life being considerably shorter.
Low battery is indicated by the appearance of the battery symbol on the display. 
If the equipment will be unused for a long period, remove the battery in order to 
prevent damage by leakage from the battery. Use only high quality brand batteries.

+

(8)

Fig. 3

Fig. 4

(16)

(15)

Fig. 5

(17)

Attaching the carrying strap and the protective caps

The carrying strap 

(Fig. 4/15)

 and the eyepiece cap 

(Fig. 4/16)

 are attached as 

shown in the illustrations.

Note:

 Feed the carrying strap only once through the eyelet on the eyepiece cap.

Depending on personal taste, use the carrying strap to connect the eyepiece cap 
either on both sides or on one side only. The eyepiece cap is held onto the eye cups
by a catch.
Before using the binoculars, remove the eyepiece cap with the index finger.
After observation, replace the eyepiece cap in order to protect the eyepieces.
The protective lens covers 

(Fig. 5/17)

 are fitted on the binoculars as shown.

TECHNICAL DATA 

8x42

10x42

8x54

10x54

Magnification

8

10

8

10

Effective lens diameter

mm

42

42

54

54

Exit pupil diameter

mm

5.3

4.2

6.8

5.4

Twilight factor

18.3

20.5

20.8

23.2

Field of view

m/1000m (ft/1000yds)

135 (405)

115 (345)

120 (360)

110 (330)

Subjective angle of view

°

62

66

55

63

Close-up setting limit

m (ft)

2.5 (8.2)

3.5 (11.5)

Diopter adjustment range

dpt

+/- 3

+/- 3

Eye relief

mm

17 (0.7)

14 (0.6)

Pupil Distance

mm

53.5 - 76

58.5 - 76

Lens type

FL

FL

Prism system

Abbe-König

Abbe-König

Coating

T*/LotuTec

®

T*/LotuTec

®

Nitrogen filling

Waterproof

mbar

400

400

Operating temperature

1

°C (°F)

-25 / +63 (-13 / +145)

-25 / +63 (-13 / +145)

Length

mm (in)

166 (6.5)

166 (6.5)

195 (7.7)

195 (7.7)

Width with an eye spacing of 65 mm

mm (in)

121 (4.8)

121 (4.8)

136 (5.4)

136 (5.4)

Weight

g (oz)

895 (31.6)

915 (32.3)

1,095 (38.6)

1,115 (39.3)

Measuring range

2

m (yds)

15 - 2,300 (16 - 2,500)

15 - 2,300 (16 - 2,500)

Measuring precision

mrad

± 1 – 600 / ± 0.5 % > 600

± 1 – 600 / ± 0.5 % > 600

Measuring time

Sek. (sec.)

< 0.3 

< 0.3 

Laser shaft length

nm

905

905

Laser beam divergence

mrad

1.6 x 0.5

1.6 x 0.5

Battery

1 x 3V / CR2

1 x 3V / CR2

Battery life at +20°C

> 2,500x

> 2,500x

Subject to changes in design and scope of supply due to technical improvements.

1

  Battery is subject to a lower temperature limit. 

Product also functions under – 10 °C /14 °F. 

2

  The range is influenced by the size and the degree of reflection of the object  

as well as by the weather and sun beams.

Summary of Contents for VICTORY RF Series

Page 1: ...vered by one or more of the following United States patents US6542302 US6816310 US6906862 For further United States patents which may cover this product see www zeiss com sports optics us patents PRECISION RANGEFINDERS FOR YOUR HUNTING AND SHOOTING ADVENTURES 8x54 10x54 8x42 10x42 ...

Page 2: ...llustration purposes only 2 VICTORY RF OWNER S MANUAL NORTH AMERICA ZEISS Sports Optics is not responsible for typographical errors Product images are for illustration purposes only Specifications and products may change without prior written notice All efforts have been made to make sure content is accurate at time of printing IDENTIFICATION OF THE COMPONENTS 1 Eye cup 2 Central focusing 3 Diopte...

Page 3: ... Never leave the binoculars in the sun for extended periods of time without the protective lens cap in place The objective lens and eyepiece can function as a burning glass and damage the interior components DANGER OF SWALLOWING Caution Do not leave the batteries and removable exterior parts within reach of children danger of swallowing Further information and safety instructions can be found in t...

Page 4: ... and debris have not contaminated the battery compartment or threads of the compartment and battery cover align and tighten the cover by turning it clockwise taking care not to cross the threads Turn the cover until it stops in order to ensure a snug fit and waterproofness NOTE At 68 F a new battery will last for over 2 500 measurements However depending on the conditions of use such as low temper...

Page 5: ...ue to secure the eyecup to the ocular housing The eyecups are ready for regular adjustments at this time Focusing the aiming mark and diopter compensation Use the range finder button Fig 1 6 to switch on the aiming mark Fig 2 9 and hold down the button Focus the aiming mark and the display by turning the right hand diopter compensation Fig 1 4 to the left or the right Then carefully use Fig 8 2 4 ...

Page 6: ...nu options by pressing the SET button once for each option NOTE See Options Table 1 BRIGHTNESS There are 11 brightness settings in the ZEISS Victory RF They are located in Menu 1 press the SET button Fig 1 7 for 2 seconds By pressing the range finding button Fig 1 6 you can select one of the 11 different brightness settings Setting 1 darkest setting possibility Setting 11 brightest setting possibi...

Page 7: ...ation System B I S II is not to be used as a replacement for the hunter s shooter s assessment of the actual shooting environment but rather to support and increase safe shooting ZEISS recommends shooting exercises from different ranges which can be used to check the correct match of the specifications with the actual points of impact also referred to as trajectory validation TABLE 2 Relates to th...

Page 8: ...u can always display the saved settings in the menu Unit Choose the appropriate unit in Menu 4 press the SET button for 2 seconds press the SET button three times Measuring Mode You can select the appropriate measuring mode for you in menu 5 press the SET button for 2 seconds press the SET button four times You can select whether you are shown the best measurement tA b or the furthest measurement ...

Page 9: ...rt setup and syncing of your data for a positive field use experience c While the VRF menus can be accessed and manipulated as an on board option the ZHA provides a faster easier and friendlier platform to input and update relevant data and settings into the VRF model ESTABLISHING A CONNECTION BETWEEN THE ZEISS HUNTING APP AND ZEISS VICTORY RF ESTABLISHING A CONNECTION BETWEEN THE ZEISS HUNTING AP...

Page 10: ...for input NOTE ZEISS strongly suggests the validation of muzzle velocity via a quality chronograph Muzzle velocity listed on some ammunition is for a specific barrel length within a controlled environment and does not factor in the many variables that are associated with obtaining true muzzle velocity through your rifle 3 Once the ZEISS Hunting App is downloaded you will need to pair the mobile de...

Page 11: ...t which you select the display data that you want to see while in the field The VRF will display these three pieces of data in the order you have selected for input Fig 15 1 The first piece of data defaults to RANGE 2 You can then select the other data from the menu options Many people like to select one of the various Holdover options ZEISS strongly suggests choosing the option that YOU are comfo...

Page 12: ...ected ensure the load is highlighted with a blue star as illustrated Fig 17 NOTE The VRF model does not have to be connected to the ZHA in the field in order to work The VRF operates as a stand alone unit ESTABLISHING DEVICE SETTINGS 1 From the ZHA Dashboard select the Connected Products tab Notice the ZEISS Victory RF model with serial number Then select Device Settings Fig 16 a From here you can...

Page 13: ...ERS ARE IN PLACE FIRST These types of cleaning solvents can and will destroy the fine and precision multi layer lens coatings Cleaning Your Lenses ZEISS recommends using original ZEISS branded lens cleaning solutions supplies and complete cleaning kits to care for the lenses on your VRF First permit heavy or large debris on the lens surfaces to fall away from the surface Try to carefully remove lo...

Page 14: ...ts Please send to ZEISS Authorized Distributor Gentec 90 Royal Crest Court Markham Ontario CANADA L3R 9X6 P 905 513 7733 STOP Before sending a product in for service please call ZEISS Sports Optics Customer Care team at the number below in order to determine if the issue concern can be resolved without having to return the product 1 800 441 3005 repair sportsoptics zeiss com Please complete the on...

Page 15: ... download the ZEISS Hunting App This free app offers useful unique and easy to use information at your fingertips Hunt longer Hunt better Hunt with confidence ZEISS 01 2020 SAP 000000 2285 607 Manual Victory RF Bluetooth Enabled Customer Service 1 800 441 3005 info sportsoptics us zeiss com Carl Zeiss SBE LLC Sports Optics Division zeiss com us sportsoptics JOIN THE CONVERSATION ...

Reviews: