![background image](http://html1.mh-extra.com/html/zeiss/sl-120/sl-120_documentation_3313162060.webp)
Electromagnetic compatibility
000000-1490-499-GA-GB-090118
54
Phenomenon
Standard
Frequency
band [MHz]
Radio service
Test level
[V/m]
Immunity to radiated
radio frequencies,
caused by wireless
communications
equipment, in
accordance with
IEC 60601-1-2:2014,
table 9
IEC 61000-4-3
380 - 390
TETRA 400
27
430 - 470
GMRS 460, FRS 460
28
704 - 787
LTE Band 13.17
9
800 - 960
GSM 800/900; TETRA 800;
iDEN 820; CDMA 850;
LTE Band 5
28
1700 - 1990
GSM 1800; CDMA 1900;
GSM 1900; DECT;
LTE Band 1, 3, 4, 25; UMTS
28
2400 - 2570
Bluetooth; WLAN 802.11b/g/n;
RFID 2450; LTE Band 7
28
5100 - 5800
WLAN 802.11 a/n
9