Maintenance and care
000000-1490-499-GA-GB-090118
36
•
Slide lamp holder (
Fig. 13) onto the contact pins of the bulb
(
•
Let the catch (
3
, Fig. 13) engage again.
•
Finally, attach the cover to the lamp housing and fix it by means of the
knurled screw (
4
Maintenance
Care and cleaning
WARNING - RISK OF ELECTRIC SHOCK
Prevent moisture from penetrating the instrument. Disconnect the power
cable from the power supply before cleaning or disinfecting the device.
CAUTION - RISK DUE TO CROSS-CONTAMINATION
Parts with which the patient has come into contact during the examination
(chin rest, forehead support) should be cleaned with a disinfectant
approved for the purpose. These parts are resistant to cleaning agents
categorized as "low" (e.g. suds, quaternary ammonium compounds) and
"intermediate" (e.g. alcohol, Javel water, iodine); classification pursuant to:
Disinfectants and activity spectrum according to the Center for Disease
Control and Prevention, Atlanta, USA.
CAUTION - RISK OF FALSE DIAGNOSIS
Use the supplied dust cover to protect the device from dust when not in
use.
CAUTION - PROPERTY DAMAGE
The national disinfecting regulations must be observed in the choice of
disinfectants and methods. Please note that some cleaning agents and
disinfectants may have an adverse effect on plastic components. Damage
caused by such disinfectants is not covered by our warranty. The surfaces of
the instrument have been tested to withstand frequent and long-term
treatment with alcoholic disinfectants and cleaning agents.
Never use aggressive or abrasive cleaning agents.