Installation
000000-1490-499-GA-GB-090118
20
Installation
Notes on installation and use
WARNING - GENERAL HAZARDS
Do not store or operate the device in ambient conditions other than those
on page 49 and following).
The device should be set up so that the power cables can be disconnected
from the power supply quickly and easily without any tools.
WARNING - RISK OF ELECTRIC SHOCK
Do not use additional extension cables or portable multiple sockets.
The electrical installation must conform to IEC 60364-7-710 or the
applicable national regulations. This includes the integration of a ground
fault circuit interrupter (GFCI).
To avert the risk of an electric shock, this device may only be connected to
a power supply with a protective ground conductor.
Ensure that the power supply plug is suitable and certified for the local
connection. If the supplied power cable must be replaced, the following
specifications must be adhered to as a minimum:
•
Protective ground conductor resistance maximum 0.1 Ohm
•
Local certification of the power cable for connection to medical devices
•
Device plug C13 conforming to IEC 60320
•
Cross-section at least 0.75 mm²/AWG 18
Hospital Grade design for specific countries (e.g. USA, Canada)
(For cables > 2.5 m the cross-section must be increased to 1.5 mm²)
The NAG SL 6 V is to be connected to mains only through a series-
connected circuit breaker 6A type B.
Installation of the device with a power supply unit which has not been
approved by Carl Zeiss Meditec must conform to IEC 60601-1 or relevant
national requirements. Responsibility for the electrical safety of the device,
including its electromagnetic compatibility (EMC), lies then exclusively with
the user.
WARNING - FIRE HAZARD
The device is not suitable for operation in explosion risk areas (e.g.
combustible mixture of anesthetic, cleaning or disinfecting agents with air,
oxygen or nitrous oxide).
The electrical installation must conform to IEC 60364-7-710.
The data concerning power consumption on the type plate must be taken
into account in the selection of overcurrent protection.