Operation of the instrument
000000-1490-499-GA-GB-090118
28
Preparations
Slit lamp
•
Set interpupillary distance of binocular tube.
•
Adjust the eyepieces:
If you wear glasses, push the eyecups in and set the diopter setting ring
to 0 D.
Other operators should pull the eyecups out and set their refractive
powers at the diopter scale of the eyepieces.
•
If necessary, factor in any instrument myopia:
The instrument myopia can be compensated using the focusing rod and
the eyepieces.
•
Adjusting the position of the diffusor:
Normally, keep the diffusor in the bottom position. Flip it up only, if you
need diffuse illumination for external examination of the eye.
1
Diffusor
E
End position
Fig. 10
Possible positions of the diffusor on the prism head
CAUTION - GENERAL HAZARDS
Before moving the slit lamp towards the patient, make sure that the diffusor
(
1
, Fig. 10) on the prism head is in one of its end positions (
E
, Fig. 10) to
avoid any eye injuries.