Safety measures
S7 / OPMI PROergo
Version 11.0
Page 16
G-30-1435-en
Before every use
•
When mounting accessory equipment, please make sure that the
admissible total weight of the surgical microscope is not exceeded.
(see "maximum load" label or the section "Technical data").
•
Before every use, make certain to compensate for any added weight; this
will enable the surgical microscope to maintain its balance in all positions
of the working range.
•
Before every use, limit the stroke of the suspension arm to prevent any
contact with the patient if the surgical microscope is lowered accidentally.
•
Check the surgical microscope for sufficient freedom of movement. The
device itself or accessories may be damaged if conflicting with each other.
•
To make sure that accessories do not fall down, check before each use
that they are firmly seated and the securing screws are firmly tightened.
•
When connecting accessories, make sure that there is sufficient clearance
and that the surgical microscope does not touch the patient.
•
Before every use, check the user settings in the software to avoid
unexpected behavior of the system.
•
Always route cables so that they do not hinder the user.
•
Never connect any unknown or defective devices to the electrical
interfaces (e.g. for data, video or foot control panel).
•
Connect components only to properly grounded outlets.
•
When setting up your system, make sure to comply with the requirements
of IEC 60601-1, chapter 16 (in some countries IEC 60601-1-1 may still be
valid).
•
Reattach any covers or caps that have been removed. Close any existing
openings with the corresponding caps.
Before starting the system, remove the protective cover used for
transportation from the objective lens.
•
If the xenon lamp is used beyond its maximum service life of 500 hours,
a sudden failure may occur. Change to the backup lamp (see Page 204)
or replace the lamp module in a timely manner (see Page 222). Reset the
service hour counter of the xenon illumination.
Summary of Contents for OPMI PROergo S7
Page 1: ...ZEISS S7 OPMI PROergo Instructions for Use G 30 1435 en Version 11 0 2018 11 30...
Page 3: ...Version 11 0 G 30 1435 en Page 3 S7 OPMI PROergo be prevented...
Page 4: ...S7 OPMI PROergo Version 11 0 Page 4 G 30 1435 en...
Page 6: ...S7 OPMI PROergo Version 11 0 Page 6 G 30 1435 en...
Page 25: ...Version 11 0 G 30 1435 en Page 25 S7 OPMI PROergo Safety measures 3 1 2 2...
Page 27: ...Version 11 0 G 30 1435 en Page 27 S7 OPMI PROergo Safety measures...
Page 29: ...Version 11 0 G 30 1435 en Page 29 S7 OPMI PROergo Safety measures 1 2 6 3 4 1 7 8 2 5...
Page 33: ...Version 11 0 G 30 1435 en Page 33 S7 OPMI PROergo Safety measures 5 3 1 1 1 1 2 6 4...
Page 35: ...Version 11 0 G 30 1435 en Page 35 S7 OPMI PROergo Safety measures 8 10 7 9...
Page 37: ...Version 11 0 G 30 1435 en Page 37 S7 OPMI PROergo Safety measures 11 13 12...
Page 39: ...Version 11 0 G 30 1435 en Page 39 S7 OPMI PROergo Safety measures 1...
Page 41: ...Version 11 0 G 30 1435 en Page 41 S7 OPMI PROergo Safety measures 1 1 5 2 1 3 4 7 9 8 5 6...
Page 45: ...Version 11 0 G 30 1435 en Page 45 S7 OPMI PROergo Safety measures 1 2...
Page 47: ...Version 11 0 G 30 1435 en Page 47 S7 OPMI PROergo Safety measures 1 1 5 2 1 3 4 7 9 8 5 6...
Page 53: ...Version 11 0 G 30 1435 en Page 53 S7 OPMI PROergo Safety measures 1 2 3 4 5 6...
Page 56: ...Safety measures S7 OPMI PROergo Version 11 0 Page 56 G 30 1435 en...
Page 59: ...Version 11 0 G 30 1435 en Page 59 S7 OPMI PROergo Device overview...
Page 63: ...Version 11 0 G 30 1435 en Page 63 S7 OPMI PROergo Device overview 2 6 4 1 5 6 3 6 6 6...
Page 65: ...Version 11 0 G 30 1435 en Page 65 S7 OPMI PROergo Device overview 4 5 6 2 1 2 3...
Page 67: ...Version 11 0 G 30 1435 en Page 67 S7 OPMI PROergo Device overview 1 3 4 2 5 6 1 2 2 1 7...
Page 69: ...Version 11 0 G 30 1435 en Page 69 S7 OPMI PROergo Device overview 10 13 11 12 9 8...
Page 75: ...Version 11 0 G 30 1435 en Page 75 S7 OPMI PROergo Device overview 1 2 3...
Page 77: ...Version 11 0 G 30 1435 en Page 77 S7 OPMI PROergo Device overview 1 2 3 4 5 6 7 8 9...
Page 79: ...Version 11 0 G 30 1435 en Page 79 S7 OPMI PROergo Device overview 1 2 3 4...
Page 81: ...Version 11 0 G 30 1435 en Page 81 S7 OPMI PROergo Device overview 1 2 3 4 5 6 7 8 9...
Page 85: ...Version 11 0 G 30 1435 en Page 85 S7 OPMI PROergo Device overview 1 3 4 2 1...
Page 87: ...Version 11 0 G 30 1435 en Page 87 S7 OPMI PROergo Device overview 1 2 3 4 5 7 6...
Page 89: ...Version 11 0 G 30 1435 en Page 89 S7 OPMI PROergo Device overview 1 2 3 4 5 6 7 8...
Page 91: ...Version 11 0 G 30 1435 en Page 91 S7 OPMI PROergo Device overview 1 4 2 3 6 5 7 8 9...
Page 93: ...Version 11 0 G 30 1435 en Page 93 S7 OPMI PROergo Device overview 1 2 3 4 5 7 6...
Page 95: ...Version 11 0 G 30 1435 en Page 95 S7 OPMI PROergo Device overview 1 2...
Page 97: ...Version 11 0 G 30 1435 en Page 97 S7 OPMI PROergo Device overview 1 2 3 4 5 6 7 8...
Page 101: ...Version 11 0 G 30 1435 en Page 101 S7 OPMI PROergo Device overview 1 2 3 4 5 6...
Page 103: ...Version 11 0 G 30 1435 en Page 103 S7 OPMI PROergo Device overview 1 2 3 4 5 6 7 8...
Page 105: ...Version 11 0 G 30 1435 en Page 105 S7 OPMI PROergo Device overview 1 4 2 3 6 5 7 8 9...
Page 107: ...Version 11 0 G 30 1435 en Page 107 S7 OPMI PROergo Device overview 1 2...
Page 109: ...Version 11 0 G 30 1435 en Page 109 S7 OPMI PROergo Device overview 2 1...
Page 111: ...Version 11 0 G 30 1435 en Page 111 S7 OPMI PROergo Device overview 1 2 3 4 5 6...
Page 113: ...Version 11 0 G 30 1435 en Page 113 S7 OPMI PROergo Device overview 2 1 6 7 5...
Page 115: ...Version 11 0 G 30 1435 en Page 115 S7 OPMI PROergo Device overview 2 4 3 6 3 7 5...
Page 133: ...Version 11 0 G 30 1435 en Page 133 S7 OPMI PROergo Device overview...
Page 137: ...Version 11 0 G 30 1435 en Page 137 S7 OPMI PROergo Device overview 1 3 2...
Page 143: ...Version 11 0 G 30 1435 en Page 143 S7 OPMI PROergo Preparations for use 1 3 4 5 6 7 2...
Page 147: ...Version 11 0 G 30 1435 en Page 147 S7 OPMI PROergo Preparations for use 3 1 2 4 5...
Page 149: ...Version 11 0 G 30 1435 en Page 149 S7 OPMI PROergo Preparations for use 1 2...
Page 151: ...Version 11 0 G 30 1435 en Page 151 S7 OPMI PROergo Preparations for use 1 2...
Page 153: ...Version 11 0 G 30 1435 en Page 153 S7 OPMI PROergo Preparations for use 1 2...
Page 159: ...Version 11 0 G 30 1435 en Page 159 S7 OPMI PROergo Preparations for use...
Page 165: ...Version 11 0 G 30 1435 en Page 165 S7 OPMI PROergo Preparations for use 1 2...
Page 171: ...Version 11 0 G 30 1435 en Page 171 S7 OPMI PROergo Preparations for use 4 1 2 1 2 2 1 3 5...
Page 178: ...Preparations for use S7 OPMI PROergo Version 11 0 Page 178 G 30 1435 en...
Page 198: ...What to do in the event of malfunctions S7 OPMI PROergo Version 11 0 Page 198 G 30 1435 en...
Page 219: ...Version 11 0 G 30 1435 en Page 219 S7 OPMI PROergo Care and maintenance...
Page 221: ...Version 11 0 G 30 1435 en Page 221 S7 OPMI PROergo Care and maintenance 1 2 5 6 4 7 3...
Page 226: ...Care and maintenance S7 OPMI PROergo Version 11 0 Page 226 G 30 1435 en...
Page 258: ...Device data S7 OPMI PROergo Version 11 0 Page 258 G 30 1435 en...
Page 266: ...Keyword index S7 OPMI PROergo Version 11 0 Page 266 G 30 1435 en...
Page 267: ...Version 11 0 G 30 1435 en Page 267 S7 OPMI PROergo Blank page for your notes...