Failure of the focusing function
Failure of the focusing function while all other functions of the sus-
pension system work correctly
•
Adjust the working distance manually using focusing knob (1).
Failure of the focusing drive
In the event of a malfunction of the focusing drive (e.g. focusing drive
moves constantly into an end position):
•
Set the mode selector switch on the connection panel to emergency
operation. This ensures that the light source continues to be operative
without any restrictions.
•
Adjust the working distance manually using focusing knob (1).
•
Continue surgery by manually operating the suspension system and
the surgical microscope by overcoming the locking effect of the
brakes.