Failure of the zoom function
Failure of the zoom function while all other functions of the suspen-
sion system work correctly
•
Adjust the magnification manually using zoom knob (1).
Failure of the zoom drive
In the event of a failure of the zoom drive (e.g. zoom drive moves con-
stantly into an end position):
•
Set the mode selector switch on the connection panel to emergency
operation. This ensures that the light source continues to be operative
without any restrictions.
•
Adjust the magnification manually using zoom knob (1).
•
Continue surgery by manually operating the suspension system and
the surgical microscope by overcoming the locking effect of the
brakes.