42
Description
G-30-1416-en
OPMI® 1FR XY Surgical Microscope on S7 Floor Stand
Issue 4.0
Printed on 14. 11. 2007
Controls and connections
Suspension arm
1
Lamp housing with halogen lamp
The stand is equipped with an illumination system for light guides. The
lamp housing contains a backup lamp which is automatically swung
into the illumination beam path when the first lamp fails.
2
Friction adjustment
Knob for adjusting the friction of the surgical microscope's rotation.
3
Coupling
for easy mounting of the surgical microscope.
4
Friction adjustment
Knob for adjusting the friction of the suspension arm's upward/down-
ward movement.
5
Locking lever for limiting the arm's downward movement
You can limit the downward movement of the suspension arm by ad-
justing a stop.
Adjustment range: from the horizontal position of the suspension arm
to its lower stop.
6
Balance setting
Knob for setting the spring force for balancing. After mounting the sur-
gical microscope including all accessories, balance the suspension
arm using this screw. The balance setting procedure is described in
the chapter “Preparations/Adjusting the system“.
-
Balance
+