88
Maintenance / Further information
G-30-1416-en
OPMI® 1FR XY Surgical Microscope on S7 Floor Stand
Issue 4.0
Printed on 14. 11. 2007
Foot control panel
Insufficient surgical field illu-
mination.
Brightness level set too low. Adjust brightness on the
suspension system's dis-
play panel or using the foot
control panel.
Halogen lamp not properly
plugged into lamp mount.
Properly push halogen lamp
into lamp mount.
Defective light guide (illumi-
nation not uniform).
Contact service dept.
Light guide probably needs
to be changed.
-
Motorized focusing system
of microscope inoperative.
Failure of suspension
system electronics.
Using the suspension arm,
set correct focal plane.
Contact service dept.
-
Motion of suspension arm is
too stiff.
Friction adjustment knob
tightened too firmly.
Slightly loosen friction ad-
justment knob.
Stand wobbles.
Floor not level. Stand base
not appropriately posi-
tioned.
Slightly change the stand
base position.
–
Problem
Possible cause
Remedy
See
Problem
Possible cause
Remedy
See
No function of foot control
panel.
Connectors on suspension
system not plugged in.
Plug in connectors.