Start-Up
N HBO 103 and N XBO 75
Carl Zeiss
HBO 103 Lamp Insertion/Change
Microscope Illuminating Systems
10
B 40-065 e 08/99
3
Start-Up
3.1
N HBO 103 Microscope Lamp
Lamp Insertion/Change
Fig. 5
Insert HBO 103 lamp
−
Before inserting the lamp, make sure that
no
connection to the line is available.
−
The lamps may only be changed when they
are cooled down; allow the microscope
illuminator to cool down for 10...15 minutes
to avoid the
risks of burns and explosion!
−
The lamp may only be removed from the
packaging and inserted in the microscope
illuminator if a protection mask and safety
gloves are worn
(see chapter 1.3).
−
All clamping adjustments must be performed
carefully; pronounced heat during operation
may result in loose contacts.
1
Fixation screw for lamp housing
2
Control wheel for collector
3
Cooling body
4
Fixation screw for lamp mount in cooling body
5
Spring lever
6
Lamp mount
7
Lamp