background image

Safety

N HBO 103 and N XBO 75

Carl Zeiss

General, EMC

Microscope Illuminating Systems

4

B 40-065 e   08/99

1

Notes on Safety

1.1

General

Please make yourself familiar with the contents
of these operating instructions before attaching
and starting up the N HBO 103 and N XBO 75
microscope illuminators. Additional information
is available from our maintenance service or
from authorized agencies.

The N HBO 103 and N XBO 75 microscope
illuminators should be used with Zeiss
microscopes exclusively.   

The N HBO 103 and N XBO 75 microscope
illuminators may only be handled by trained
personnel.

To ensure that the illuminators function
properly with the microscope, the appropriate
microscope manual must be observed.

The manuals
"

ebq 100 dc

 

– Electronic transformer  for the N

HBO 103 microscope illuminator“

 and

"

ebx 75 isolated – Electronic transformer for

the N XBO 75 microscope illuminator“

must be observed.
Special regard must be paid to the notes on
instrument safety included in this manual.

The data sheets from the lamp manufacturers
must be observed when the HBO 103 and XBO
75 lamps are operated.

To guarantee the safe operation and function
of the N HBO 103 and N XBO 75 microscope
illuminators, it is necessary under any
circumstances to take the precaution measures
and observe the warnings contained in the
operating instructions.

The following symbols are used for the
warnings:

CAUTION!

Non-observance of the safety notes constitutes
a hazard for the user.

CAUTION!

Dangerous electrical voltage

CAUTION!

Disconnect the instrument from the line before
opening it!

CAUTION!

Hot surface

CAUTION!

Non-observance of the safety notes constitutes
the risk of damage to the instrument.

NOTE!

Notes which must be observed when using the
N HBO 103 and N XBO 75 microscope
illuminators.

1.2

Regulations concerning
Instrument Safety and
Electromagnetic
Compatibility (EMC)

The N HBO 103 and N XBO 75 microscope
illuminators have been designed and tested in
compliance with EN 61010, Part 1 (VDE 0411).
From the safety viewpoint, they left the factory
in a perfect state.
They meet the requirements of EC directives
73/23 and 89/336 and the EMC legislation,
version of Nov.9, 1992.
Radio-screened in accordance with EN 55011,
Class B, insensitivity to noise in accordance with
EN 50082-2.
Compliance with the EC directives mentioned is
documented by the CE-certificate.

Summary of Contents for N HBO 103

Page 1: ...Operating Manual N HBO 103 and N XBO 75 Microscope Illuminating Systems ...

Page 2: ...t The specifications are subject to change Unless expressly authorized forwarding and duplication of this document and the utilization and communication of its contents are not permitted Violations will entail an obligation to pay compensation All rights reserved in the event of granting of patents or registration of a utility model Published by Carl Zeiss Microscopy D 07740 Jena Phone 49 36 41 64...

Page 3: ...Lamp 15 3 5 Focus and Center the HBO 103 XBO 75 Lamps in the Microscope Lamp Housing 16 3 6 Change the Collector 17 Illustrations Fig 1 Overall view of the N HBO 103 microscope illuminating system 7 Fig 2 Overall view of the N XBO 75 microscope illuminating system 7 Fig 3 Line spectrum of the HBO 100 103 mercury short arc lamps 9 Fig 4 Line spectrum of the XBO 75 xenon short arc lamp 9 Fig 5 Inser...

Page 4: ...BO 75 lamps are operated To guarantee the safe operation and function of the N HBO 103 and N XBO 75 microscope illuminators it is necessary under any circumstances to take the precaution measures and observe the warnings contained in the operating instructions The following symbols are used for the warnings CAUTION Non observance of the safety notes constitutes a hazard for the user CAUTION Danger...

Page 5: ...temperatures must be observed in the same way as those of the microscope stands 1 4 Hints on Use The lamps especially the XBO 75 are under high pressure even if they are not switched on They may only be inserted in the microscope illuminator if a safety mask and safety gloves see chapter 1 3 are worn The safety regulations of the lamp manufacturer must be observed when the XBO 75 is inserted The H...

Page 6: ...The lamps must be exchanged as described in chapters 3 1 3 2 Before each lamp exchange make sure to allow the N HBO 103 or N XBO 75 microscope illuminator to cool down sufficiently long in order to reduce the pressure wait at least for 15 minutes Always disconnect the instrument from the line before each lamp exchange or any other work inside the instrument The safety regulations of the lamp manuf...

Page 7: ...attachment to the microscope reflector double lens collector or quartz collector for UV light heat protection filter HBO 103 W 2 mercury vapor short arc lamp or XBO 75 W 2 xenon short arc lamp transformer ebq 100 dc for N HBO 103 or transformer ebx 75 isolated for N XBO 75 line cable with earth contact plug or line cable with American flat plug HBO 103 XBO 75 Microscope Illuminating Systems Lamp N...

Page 8: ...000 1003 928 Transformer N for XBO 75 Cat No 000000 1003 924 Line connection 90 250 V AC 90 250 V AC Line frequency 50 60 Hz 50 60 Hz Power consumption 265 VA 130 VA Fuse F1 F2 T 3 15 A F1 F2 T 2 A Lamp Mercury vapor short arc lamp Xenon short arc lamp For the technical data of the lamps please see the information provided by the manufacturer Recommended OSRAM Type HBO 103 W 2 Cat No 380301 9350 0...

Page 9: ...N XBO 75 Technical Data Microscope Illuminating Systems Line spectra Carl Zeiss B 40 065 e 08 99 9 Fig 3 Line spectrum of the HBO 100 103 mercury short arc lamps Fig 4 Line spectrum of the XBO 75 xenon short arc lamp ...

Page 10: ...d down allow the microscope illuminator to cool down for 10 15 minutes to avoid the risks of burns and explosion The lamp may only be removed from the packaging and inserted in the microscope illuminator if a protection mask and safety gloves are worn see chapter 1 3 All clamping adjustments must be performed carefully pronounced heat during operation may result in loose contacts 1 Fixation screw ...

Page 11: ...the mount 5 6 in the lamp base Loosen fixation screw 5 4 on the cooling body remove defective lamp and dispose of it in accordance with the appropriate regulations see section 1 6 Notes on disposal For initial start up remove the safety appliance first Insert cathode smaller base of the new lamp into the cooling body until stop When inserting the lamp make sure to use the correct hole diameter in ...

Page 12: ...under high pressure even if it is not switched on It may only be removed from the packaging and inserted in the microscope illuminator if a protection mask and safety gloves are worn See section 1 3 The safety regulations of the lamp manufacturer must be observed when the lamp is inserted or changed See section 1 3 Notes on use All clamping adjustments must be performed carefully pronounced heat d...

Page 13: ... 6 4 on the cooling body remove defective lamp and dispose of it in accordance with the appropriate regulations see section 1 6 Notes on disposal For initial start up remove the safety appliance first Insert anode larger base of the lamp in the cooling body until stop When inserting the lamp make sure to use the correct hole diameter in the cooling body the mount in the cooling body is marked with...

Page 14: ...escribed in the supplied ebq 100 dc operating instructions Figure 7 Connect transformer to the line Switch on microscope lamp housing via line switch at the front of the transformer the lamp will ignite automatically Figure 8 Meaning of the luminiscence diodes Power Line control in the line switch Lamp The green LED lights up when the lamp is on Temp The yellow LED indicates inadmissible heating o...

Page 15: ...ansformer to the line Switch on microscope lamp housing via line switch at the front of the transformer the lamp will ignite automatically Figure 10 Meaning of the luminiscence diodes Power Line control in the line switch SAFETY The green LED indicates whether the safety circuit is closed and whether the instrument is ready for operation Lamp The green LED lights up when the lamp is on The green L...

Page 16: ...ork while the lamp is exposed do not look directly into the lamp or its reflected light wear safety glasses sunglasses Direct microscope lamp housing against wall approx 3m away Switch on the lamp starting time 3 minutes Adjust collector via operating wheel 12 1 and produce a sharp image of the lamp arc on the wall adjust the mirror until the mirror image is in focus Fig 11 Fig 12 both images shou...

Page 17: ...heel Fig 13 Change the collector Loosen clamping ring 13 1 Remove heat protection filter 13 2 Pull out collector operating wheel 13 5 against resistance until the pin allows access to groove 13 4 Remove collector from the illuminator Push in new collector until pin of the pushed out operating wheel locks into the groove the collector can be adjusted via the operating wheel Insert heat protection f...