12
MCS 600 | User Manual
General Safety Information
MCS 600
2.7 Storage and transport
No special transport container or locking device is provided with the system.
Although the components are very robust, collision of any sort must be avoided.
Install the measuring system in a safe, easily accessible location. Please note the
specified environmental parameters when storing and working with the system.
2.8 Installation and connection requirements
The MCS 600 system is designed to be used in a horizontal position with the
housing feet facing downward. We do not recommend placing the device in any
other position, as this will impair the device's ability to dissipate heat. This also
applies to installations involving a protective housing (an optional accessory for
using the MCS 600 in environments requiring higher IP ratings).
Line voltage
100 – 240 V AC
Power frequency
50/60 Hz
Working temperature range
*exception for MCS 611 NIR 1.7 for lab applications
(000000-2187-068)
+5 °C to 35 °C*
+5 °C to 25 °C
Storage/transport temperature
– 40 °C to + 70 °C
Max. humidity
95 % non-condensing
Use environment altitude above sea level
Up to 2000 m
Summary of Contents for MCS 600
Page 1: ...User Manual Carl Zeiss Spectroscopy GmbH MCS 600...
Page 14: ...14 MCS 600 User Manual System Overview MCS 600...
Page 22: ...22 MCS 600 User Manual MCS 600...
Page 30: ...30 MCS 600 User Manual MCS 600...
Page 32: ...32 MCS 600 User Manual MCS 600...
Page 36: ...36 MCS 600 User Manual MCS 600 1 3 4 5 6 2 Fig 17 BLX 600 SMA lamp replacement...
Page 40: ...40 MCS 600 User Manual MCS 600 1 3 4 5 6 2 Fig 20 BLX 600 CZ6 lamp replacement...
Page 44: ...44 MCS 600 User Manual MCS 600 1 3 4 5 6 2 Fig 23 BLX 606 lamp replacement...
Page 48: ...48 MCS 600 User Manual MCS 600 1 3 4 6 5 7 8 9 2 Fig 26 CLD 600 lamp replacement...
Page 52: ...52 MCS 600 User Manual MCS 600 1 3 4 5 6 2 Fig 29 CLH 600 lamp replacement...
Page 56: ...56 MCS 600 User Manual MCS 600 1 3 4 5 6 2 Fig 32 CLH 600 F lamp replacement...
Page 60: ...60 MCS 600 User Manual MCS 600 1 3 4 5 2 Fig 35 CLH 606 A bulb replacement...
Page 66: ...66 MCS 600 User Manual MCS 600 Network Connection MCS 600...
Page 68: ...68 MCS 600 User Manual Maintenance and Care MCS 600...