46
MCS 600 | User Manual
MCS 600
3.11.2 Operating notes
Note
The spectral emission radiation characteristic of the deuterium lamp
will change during the warm-up phase. A minimum warm-up time
of approx. 30 minutes must therefore be observed as the emitter is
then able to achieve a stable operating condition. For very precise
measurements, a warm-up time of at least 60 minutes is necessary so
that the optical and mechanical components of the bulb can achieve a
constant working temperature.
Note
Turning the lamp on and off frequently can considerably reduce the
emitter's average life span. The lamp should therefore not be turned off
if work is halted for relatively short periods of time.
3.11.3 Safety information for CLD 600 lamp cassette
Caution
UV radiation!
The deuterium lamp emits light in the ultraviolet range! Do not look
directly into the beam path! Doing so will result in injury to the eyes.
Please observe the relevant regulations when handling UV light sources.
3.11.4 Spare parts for CLD 600 lamp cassette
Designation
Cat. no.
Deuterium lamp for CLD 600
000000-1446-440
Summary of Contents for MCS 600
Page 1: ...User Manual Carl Zeiss Spectroscopy GmbH MCS 600...
Page 14: ...14 MCS 600 User Manual System Overview MCS 600...
Page 22: ...22 MCS 600 User Manual MCS 600...
Page 30: ...30 MCS 600 User Manual MCS 600...
Page 32: ...32 MCS 600 User Manual MCS 600...
Page 36: ...36 MCS 600 User Manual MCS 600 1 3 4 5 6 2 Fig 17 BLX 600 SMA lamp replacement...
Page 40: ...40 MCS 600 User Manual MCS 600 1 3 4 5 6 2 Fig 20 BLX 600 CZ6 lamp replacement...
Page 44: ...44 MCS 600 User Manual MCS 600 1 3 4 5 6 2 Fig 23 BLX 606 lamp replacement...
Page 48: ...48 MCS 600 User Manual MCS 600 1 3 4 6 5 7 8 9 2 Fig 26 CLD 600 lamp replacement...
Page 52: ...52 MCS 600 User Manual MCS 600 1 3 4 5 6 2 Fig 29 CLH 600 lamp replacement...
Page 56: ...56 MCS 600 User Manual MCS 600 1 3 4 5 6 2 Fig 32 CLH 600 F lamp replacement...
Page 60: ...60 MCS 600 User Manual MCS 600 1 3 4 5 2 Fig 35 CLH 606 A bulb replacement...
Page 66: ...66 MCS 600 User Manual MCS 600 Network Connection MCS 600...
Page 68: ...68 MCS 600 User Manual Maintenance and Care MCS 600...