5
3. Start Lasers
Click the
Acquire
button on the
toolbar. The lower toolbar will now
change to the Acquire sub-toolbar and
show the acquisition controls.
Click the
Laser
button.
Turn on the desired lasers
position: HeNe’s to
ON
, Argon
to S
tandby.
When the
Status
reads
Ready
, click the
On
buttons.
Also turn
OFF
any lasers that
have been left on for you, but
you will not be using.
If the previous user has left the
lasers on for you, Argon and
Enterprise may already be in
standby mode and ready to be
tuned on.
When all the
Status
reads
Ready
, close the Laser Control window.