CAUTION:
Allow the lamp to cool completely
(approximately five minutes) before handling it.
CAUTION:
Do NOT touch the two disk-shaped filters.
5.
Remove the connector cable by pulling its connector straight
upward. Then use a screwdriver to loosen the screw.
Figure 4. Connector Removal
6.
Now, slide the wishbone-shaped plate that holds the lamp
assembly in place away from you.
Figure 5.