General Information
Humphrey Field Analyzer 3 Service Manual
2
2660021151225 A
Classification / Compliance
Class I Equipment –
Protection against electrical shock.
Type B –
Degree of protection against electric shock of applied
part (chin, forehead, and patient button).
Ordinary Equipment (IPXO) – Degree of protection against ingress of liquids (none).
CAUTION:
This instrument has no special measures to protect
against harmful ingress of water or other liquids (classified IPXO
– ordinary equipment).
Do not place containers of liquid on or near the instrument, and
do not use aerosols on or near it.
Continuous Operation – Mode of operation.
•
Not suitable in the presence of flammable anesthetic.
Complies with 93/42/EEC Medical Device Directive
Complies with US and Canadian medical electrical system safety requirements
System Electrical Information
Instrument Electrical Rating:
•
100 – 120V
~
, 50/60 Hz, 4.0A
•
230V
~
, 50/60 Hz, 1.8A
Power Table Electrical Rating:
•
HFA 120: 120V
~
, 60 Hz, 8.0A
•
HFA 230: 230V
~
, 50 Hz, 6.0A
Power Outlet Ratings:
•
HFA 120: 120V
~
, 5A Max.
•
HFA 230: 230V
~
, 3A Max.