Humphrey Field Analyzer 3 Service Manual
20
2660021151225 A
HFA3 Repacking Instructions
CAUTION —
The bowl is easily damaged if abused. DO NOT store
anything inside the bowl. If the instrument is received with objects
stored inside the bowl, your account will be charged for ensuing
repairs.
1) Disconnect the power cords and all peripherals from the HFA3.
2) Pack the patient button, power cord, and all other accessories in the
accessory box, if appropriate.
3) Put the trial lens holder in the down position.
4)
Rotate the HFA turret completely clockwise. To hold the turret in
place:
loop one end of the rubber-band turret restraint around the end
of the small turret pulley;
pass the rubber band behind the turret and loop the other end
around the left-hand bowl lamp baffle inside the bowl.
5) Remove the perforated cutout from the right side of the base foam
insert.
6)
Place the base foam insert inside the cardboard bottom on the pallet.
7)
Place the HFA in the base as indicated on the drawing.
8)
Remove the perforated cutout from the right side of the right top
foam insert.
9)
Orient the top right and top left foam inserts to conform with the
surface of the field analyzer contours.
10)
If you are shipping the keyboard back with the unit, place the
keyboard in the slot on the rear top of the foam inserts.
11)
Place the accessory kit on top of the left and right foam inserts.
12)
Slide cardboard tube over the left and right foam inserts until bottom
edges touch the bottom of the cardboard bottom.
13)
Place the cardboard top on the cardboard tube.
14)
Secure entire shipping container with repackaging straps and buckles.
See the detail of how to use the buckles.