Page | 8  

 

Switch on Procedure 

1)

 

Switch on all the labelled 13A mains sockets with the Micr X-
Cite socket on first.  

2)

 

All the microscope control boxes including the Andor EM-CCD camera 
should come on with the exception of the Tokai Hit incubator heater 
unit that is only switched on as required. 

3)

 

Switch on the system PC and 32” monitor and login to the WTCHG 
network. 

4)

 

Start Zen Black (don’t select the Zen Blue option as that is for off-line 
image analysis only). Start up the Elyra hardware via the software 
prompt.

 

5)

 

Place the specimen onto the microscope

 

6)

 

Use the triple filter block FSet77HE (GFP/DsRed) or BF (bright field 
transmission) to view the specimen down the microscope via the 
eyepieces (filter selection and open/close shutters options are under 
‘Locate’ in Zen Black).

 

7)

 

Use the X-Cite footswitch or Start-Stop button to illuminate the 
specimen with excitation light, in combination with open/close filter 
options in Zen Black.

 

8)

 

Ensure that the doors and viewing panels are closed on the black stage 
incubator or the lasers will switch off. Wait ~5 minutes until the Andor 
EM-CCD camera electronics have cooled down before acquiring images.

 

9)

 

Go to Acquisition in Zen Black to capture images

  

The microscope stand must not be left pushed backwards on the microscope or the 
lasers will not fire (there is a safety cut-out).

 

10)

 

Create and save filter and EM-CCD camera settings for the 
fluorochromes  in your sample. 

 

11)

 

Leave the Elrya microscope system on for at least 5 minutes before 
scanning as the Andor EM-CCD camera needs time for its electronics to 
cool down.

 

12)

 

The system has three laser lines at 405nm DAPI, 488nm GFP and 
561nm TRITC/DsRed. There are four different emission filter sets for 
DAPI, GFP, DsRed and Cy5 far red, selectable via Zen Black. 

Warning: The Elyra lasers are shuttered off by magnetic micro-switches if one of the 
access doors is left open on the blackened live cell incubator chamber.

 

Summary of Contents for ELYRA

Page 1: ...Page 1 Wellcome Trust Centre for Human Genetics Cellular Imaging Microscopy Core The Zeiss Elyra TIRF Confocal Abridged INSTRUCTIONS...

Page 2: ...super resolution with temperature control Plan Apochromat 10x NA 0 3 air objective Plan Apochromat 20x objective NA 0 8 air objective Alpha Plan Apochromat 100x NA 1 46 oil objective Andor iXon DU 89...

Page 3: ...ain with the camera it s locked in software The EM Gain going from 1 to 200 goes from the equivalent of second to a 20ms to frame rate Avoid putting too much light onto the cameras EMCCD sensor The EM...

Page 4: ...r use If you are using the heated incubator at 37oC you should switch on the incubator at least 2 hours before imaging For the most stable time lapse imaging at 37oC leave the incubator on overnight T...

Page 5: ...yra microscope The tube lens can be 1x or 1 6x always use the 1 6x with super resolution If the 1x tube lens is selected you can see the shadow of the 1 6x down the microscope vanishes when the 1 6x i...

Page 6: ...n OOO appears on the display this is the lamp intensity Press Up Down to select 0 12 25 50 and 100 lamp intensity Warning things can very bright down the eyepieces even with the standard FITC TRITC Fa...

Page 7: ...ails on the operation of the Elrya TIRF super resolution microscope see the Ziess Elyra manual and in particular the Elyra Quick Guide after chapter 7 of the manual a pdf download of the Quick Guide i...

Page 8: ...ith excitation light in combination with open close filter options in Zen Black 8 Ensure that the doors and viewing panels are closed on the black stage incubator or the lasers will switch off Wait 5...

Page 9: ...y possibility of damage during start up when the XY motorised stage initialises Note The top of the 100x objective rotates so avoid accidently turning it 4 Drop the 10x objective nearer to the Load lo...

Page 10: ...ave any microscope imaging problems or queries just ask a member of the Microscopy Core staff for help There is additional help and advice on our Core web pages http www well ox ac uk microscopy WTCHG...

Reviews: