background image

Page | 3  

 

General information 

The system can only image green GFP/FITC, red DsRed/RFP/TexsRed and blue 
DAPI dyes. 

The Andor iXon DU 897 EM-CCD camera has a maximum resolution of 512 x 
512 pixels. It captures black and white images with a 16-bit depth of 65,536 
grey levels (with 0 being pure black- and 65,535 being pure white). 

Zen Black doesn’t automatically save images – you must save them as you go 
along. 

Use ‘snap’ in Zen Blue to clear noise from the camera detectors before 
imaging.  Setting ‘snap’ for every image will slow down acquisition a lot, but 
you can just ‘snap’ clear at the beginning of an image sequence. Adjust 
exposure time, laser power and EM Gain for brightness in ‘Acquisition’. 

Unlike the Zeiss Spinning Disk microscope in OPIC, you can’t go above 300 in 
EM Gain with the camera (it’s locked in software ). The EM Gain going from 1 
to 200 goes from the equivalent of ½ second to a 20ms to frame rate. Avoid 
putting too much light onto the cameras EMCCD sensor. The EM-CDD camera 
is 16-bit (64,000 shades of grey). 

Leave the Elrya system on for 5 minutes before scanning as the Andor EM-CCD 
camera needs time for its electronics to cool down. 

The system has three laser lines at 405nm DAPI, 488nm GFP and 561nm 
TRITC/DsRed. There are four different emission filter sets for DAPI, GFP, DsRed 
and Cy5 far red, selectable via Zen Black. 

In Zen Black click Live for a faster ‘preview’ imaging with more noise.  

You can save configurations and the menu screen setup in Zen Black. You can 
reuse microscope settings from a stored image. The images are generally saved 
in Zeiss *.czi format although you can use *.lsm. To prevent PC crashes with 
Zen Black, it’s a good idea to limit the number of open images. 

Log-in to the PC via the account ‘

cyto\lsm user

’. 

 

 

Summary of Contents for ELYRA

Page 1: ...Page 1 Wellcome Trust Centre for Human Genetics Cellular Imaging Microscopy Core The Zeiss Elyra TIRF Confocal Abridged INSTRUCTIONS...

Page 2: ...super resolution with temperature control Plan Apochromat 10x NA 0 3 air objective Plan Apochromat 20x objective NA 0 8 air objective Alpha Plan Apochromat 100x NA 1 46 oil objective Andor iXon DU 89...

Page 3: ...ain with the camera it s locked in software The EM Gain going from 1 to 200 goes from the equivalent of second to a 20ms to frame rate Avoid putting too much light onto the cameras EMCCD sensor The EM...

Page 4: ...r use If you are using the heated incubator at 37oC you should switch on the incubator at least 2 hours before imaging For the most stable time lapse imaging at 37oC leave the incubator on overnight T...

Page 5: ...yra microscope The tube lens can be 1x or 1 6x always use the 1 6x with super resolution If the 1x tube lens is selected you can see the shadow of the 1 6x down the microscope vanishes when the 1 6x i...

Page 6: ...n OOO appears on the display this is the lamp intensity Press Up Down to select 0 12 25 50 and 100 lamp intensity Warning things can very bright down the eyepieces even with the standard FITC TRITC Fa...

Page 7: ...ails on the operation of the Elrya TIRF super resolution microscope see the Ziess Elyra manual and in particular the Elyra Quick Guide after chapter 7 of the manual a pdf download of the Quick Guide i...

Page 8: ...ith excitation light in combination with open close filter options in Zen Black 8 Ensure that the doors and viewing panels are closed on the black stage incubator or the lasers will switch off Wait 5...

Page 9: ...y possibility of damage during start up when the XY motorised stage initialises Note The top of the 100x objective rotates so avoid accidently turning it 4 Drop the 10x objective nearer to the Load lo...

Page 10: ...ave any microscope imaging problems or queries just ask a member of the Microscopy Core staff for help There is additional help and advice on our Core web pages http www well ox ac uk microscopy WTCHG...

Reviews: