Start-Up
Carl Zeiss
Attaching epi-fluorescence illuminator
Axiostar
plus
2-22
B 40-815 e 12/01
1
30°
ICS
Tube
2
Epi-fluorescence illuminator
3
Filter stop carrier 2
4
Filter stop carrier 1
5
Hexagonal screw (on stand)
6
Clamping screw
7
Clamping screws (back)
Fig. 2-20 Attaching epi-fluorescence illuminator
2.11
Attaching epi-fluorescence
illuminator
The epi-fluorescence illuminator can be used
only in conjunction with a binocular
tube/phototube 30°.
•
If necessary, remove the 45° tube or 30°
tupe and adapter for 30° tubes.
•
Replace knurled screws on the stage mount
with hexagonal screws (2-20/
5
).
•
Press epi-fluorescence illuminator (2-20/
2
)
at a slight angle to the right into the spring
of the stand mount.
•
Then press the epi-fluorescence illuminator
into the dovetail mount of the stand by
lowering the left side.
•
Align epi-fluorescence illuminator to the
outer contours of the stand.
•
Loosen the two back clamping screws
(2-20/
7
) on the epi-fluorescence illuminator.
•
First tighten the hexagonal screws (2-20/
5
)
and then the two back clamping screws
(2-20/
7
) to hold the illuminator in position.
•
Place 30°
ICS
tube (2-20/
1
) with ring mount
in the epi-fluorescence illuminator opening
(2-20/
2
), align to the outer edges and
tighten the clamping screws (2-20/
6
) on the
epi-fluorescence.