Instrument description
Axiostar
plus
Technical Data
Carl Zeiss
40-815 e 12/01
1-13
(7)
Opto-mechanical data
Stand with stage focusing........................................................................ with coarse drive
2
(4 mm/U)
and fine drive (0.4 mm/U)
overall lift 15 mm
Objective change ...............................................................manually via 4/5-position objective revolver
Objectives .............................................................................. ICS line of objectives with W 0.8 thread
Eyepieces ...................................................................................................... 30 mm plug-in diameter
with field number 18.................................................. PL 10
×
/18 Br. and PL 10
×
/18 Br. foc. or
with field number 20................................................E- PL 10
×
/20 Br. and E- PL 10
×
/20 Br. foc.
Specimen stage ..................................................... mechanical stage 75
×
30 R/L with ceramic surface
Dimensions (width
×
depth).............................................................................. 170
×
140 mm
Travel range (width
×
depth) ................................................................................ 75
×
30 mm
Coaxial drive ..........................................................................................right or left selectable
Vernier............................................................................................... readable from the right
Specimen holder ..............................................................................with spring clip to the left
Abbe-condenser 0.9/1.25 .............................for V
Obj.
<
4
×
with wide angle function 445312-0000-000
Binocular tube 45°/20
maximum field number .......................................................................................................20
interpupillary distance ........................................................can be set between 55 and 75 mm
viewing angle.................................................................................................................... 45°
viewing height ................................................................................................. 425
−
470 mm
visual port ......................................................................................................... tube factor 1
×
Binocular phototube 45°/20
maximum field number .......................................................................................................20
interpupillary distance ........................................................can be set between 55 and 75 mm
viewing angle.................................................................................................................... 45°
viewing height ................................................................................................. 425
−
470 mm
visual port ......................................................................................................... tube factor 1
×
camera/video port ............................................................................................ tube factor 1
×
camera/video port ..........................................................................................interface 60 mm
switched via folding prism ......................................................................100% vis / 100% doc
2
The scale on the coarse drive (0 to 400) permits the orienting measurement of the object thickness:
1 increment corresponds to approx. 5 µm