DESCRIPTION OF THE INSTRUMENT
ZEISS
Control and functional elements on microscope components
Axiolab 5
42
430037-7444-001
05/2019
2.5.8
Polarizer
Polarizer D, 90° rotatable, switchable
(Fig. 2-21/3)
−
Polarizer can be swiveled in/out using handle
(Fig. 2-21/
1
)
−
Polarizer with lever (Fig. 2-21/
2
), 90°
rotatable
Polarizer, fixed with lambda plate, rotatable,
(Fig. 2-21/7)
−
Polarizer can be swiveled in/out using handle
(Fig. 2-21/
5
)
−
Lambda plate can be swiveled in/out using
handle (Fig. 2-21/
4
)
−
Lambda plate with lever (Fig. 2-21/
6
), rotatable
ATTENTION
If the condenser carrier with mounted polarizer is moved too far down, the polarizer may
collide with the field diaphragm and damage it!
2.5.9
Nosepiece with objectives
−
Nosepiece 5x with M27 threaded insert for
five objectives.
−
You can change the objectives quickly by
turning the nosepiece on knurled ring
(Fig. 2-22/
2
).
−
Incorporates slot (Fig. 2-22/
3
) for 6x20mm
slider (compensators, analyzers, quarter plates).
−
Stand for transmitted light polarization and
stand for transmitted light conoscopy with
5_position nosepiece, of which one position
is fixed and four positions can be centered
with the aid of two screws each
(Fig. 2-22/
1
).
ATTENTION
Do not overtighten the screws (Fig. 2-22/
1
) on the stop.
Fig. 2-21
Polarizers
Fig. 2-22
Nosepiece of the transmitted/
reflected light polarization stand
with mount for compensators