To calibrate the zero set point, The output from the
transmitter should be 1.00 VDC.
1. Press & release the *Read key and the actual
voltage present from the transmitter will be shown
flashing in the display, figure 98.
2. Press & release the *Accept key to load the zero
point into the memory. This value will stop flashing
when this is accomplished, figure 99.
Note: if an error message appears, figure 100, the
voltage is outside of the calibration range of .6V-1.4VDC.
Return all valves and pressures to normal settings
before restarting the system.
Or
pmp bat vmtr sig tnk
Zero Adjustment
*Read 1.0V *Esc
Figure 98
pmp bat vmtr sig tnk
Zero Adjustment
*Accept 1.0V *Esc
pmp bat vmtr sig tnk
Zero Adjustment
ERROR 2.0V *Esc
Figure 100
Figure 99