MultiLab 4010-1W
Overview
ba76141e07
08/2018
9
have quick-lock couplings with which to fix the sensors to the meter.
Sensor data from
IDS sensors
IDS sensors transmit the following sensor data to the meter:
SENSOR ID
– Sensor name
– Sensor series number
Calibration data
Measurement settings
The calibration data are updated in the IDS sensor after each calibration proce-
dure. A message is displayed while the data are being updated in the sensor.
1.3.2
Wireless operation of IDS sensors
With the aid of the adapters in the IDS WA Kit, IDS sensors with plug head
connectors (variant P) can be wirelessly connected to your MultiLab 4010-1W.
(MultiLab 4010-1 IDS: from software version V2.00)
Two adapters, one at the IDS meter (IDS WA-M) and one at the sensor (IDS
WA-S), replace the sensor cable with an energy-saving Bluetooth LE radio
connection.
1.3.3
Automatic sensor recognition
The automatic sensor recognition for IDS sensors allows
to operate an IDS sensor with different meters without recalibrating
to assign measurement data to an IDS sensor
– Measurement datasets are always stored and output with the sensor
name and sensor serial number.
to assign calibration data to an IDS sensor
– Calibration data and calibration history are always stored and output with
the sensor name and sensor series number.
to hide menus automatically that do not concern this sensor
In the measured value display, you can display the sensor name
and series number of the selected sensor with the [Info] softkey.
You can then display further sensor data stored in the sensor with
the [More] softkey (see section 4.1.5 S
Further information on the wireless operation of IDS sensors:
Web resources
Operating manual of the IDS WA Kit.