MultiLab 4010-1W
Settings
ba76141e07
08/2018
71
Setting menu 4310
Menu item
Possible
setting
Explanation
Type
Used measuring cell
Cal
Measuring cells whose cell con-
stants are determined through
calibration in the check- and cali-
bration standard 0.01 mol/l KCl
(1413 µS/cm @25 °C).
Calibration range:
0.450 to 0.500 cm
-1
The currently valid cell constant
is displayed in the status line.
man
Freely (manually) adjustable cell
constant in the range 0.450 ...
0.500 cm
-1
.
Cell const. man
0.450 ...
0.475
...
0.500 cm
-1
Display and setting options for
the manually adjustable cell con-
stant.
This menu item is only available
when
Type
man
is set.
The current conductivity value is
displayed in the status line.
Temp. comp. (TC)
/
Method
nLF
Lin
off
Procedure for temperature com-
pensation (see section 8.2 T
57).
This setting is only available for
the measured parameters, con-
ductivity (
ϰ
) and resistivity (
ρ
).
Temp. comp. (TC)
/
Linear coeff.
0.000 ...
2.000
...
3.000 %/K
Coefficient of the linear tempera-
ture compensation.
This menu item is only available
when the linear temperature
compensation is set.
Temp. comp. (TC)
/
Reference temp.
20 °C
25 °C
Reference temperature
This setting is only available for
the measured parameters, con-
ductivity (
ϰ
) and resistivity (
ρ
).
Multiplier for TDS
0.40 ...
1.00
Multiplier for TDS value