Sondes
Section 2
YSI Incorporated
Environmental Monitoring Systems Operations Manual
2-33
If you have a depth sensor installed, you can maintain the default settings of 40 and 0 for
2-Latitude
and
3-Altitude
, respectively, without affecting your ability to learn the basic calibration and operation of the
sonde. However, if you know the appropriate values for your location, change them. When this setup is
verified, press
Esc
or
0
to return to the Advanced menu. For more information, see
Section 2.9.8,
Advanced
.
The display under
3-Sensor
may be different from the one shown in the example above, depending on the
sensors that are installed in your unit. For example, if you do not have a chlorophyll probe, the last two
entries (which are relevant only to chlorophyll) will not appear.
When this setup is verified, press
Esc
or
0
to return to the Advanced menu. For a detailed explanation of
the choices in the Advanced menu, see
Section 2.9.8, Advanced
. Press
Esc
or
0
to back up to the Main
menu.
The sonde software is now set up and ready to calibrate and run.
------------Advanced sensor-----------
1-TDS constant=0.65
2-Latitude=40
3-Altitude Ft=0
4-(*)Fixed probe
5-( )Moving probe
6-DO temp co %/C=1.1
7-DO warm up sec=40
8-( )Wait for DO
9-Wipes=1
A-Wipe int=5
B-SDI12-M/wipe=1
C-Turb temp co %/C=0.3
D-(*)Turb spike filter
E-Chl temp co %/C=0
F-( )Chl spike filter
Select option (0 for previous menu):
------------------Main-----------------
1-Run 5-System
2-Calibrate 6-Report
3-File 7-Sensor
4-Status 8-Advanced
Select option (0 for previous menu):