CHECKING ADVANCED SETTINGS
Select
Advanced
from the Main menu. The following menu will be displayed.
Select
Setup
from the Advanced menu.
Make sure that, other than
Auto sleep RS232
, all entries are activated or deactivated as shown above.
For sondes which will be used in sampling studies where the user is present and observes readings in real-
time,
Auto sleep RS232
should usually be “off
‟
. For sondes that will be used in unattended monitoring
studies,
Auto sleep RS232
should usually be “on”. This is described in detail in
Section 2.9, Sonde Menu
.
When this setup is verified, press
Esc
or
0
to return to the Advanced menu.
Select
3-Sensor
from the Advanced menu and make certain that the entries are identical to those shown
below.
----------------Advanced--------------
1-Cal constants
2-Setup
3-Sensor
4-Data filter
Select option (0 for previous menu):
-------------Advanced setup-----------
1-(*)VT100 emulation
2-( )Power up to Menu
3-( )Power up to Run
4-( )Comma radix
5-(*)Auto sleep RS232
6-(*)Auto sleep SDI12
7-( )Multi SDI12
8-( )Full SDI12
Select option (0 for previous menu): 0