background image

 

YSI  550A

 

Y S I

Environmental

 

Page  5 

 

PRINCIPLES OF OPERATION 

 
The sensor consists of a silver body as the anode and a circular gold cathode embedded in the end. In 
operation, this end of the sensor is filled with a solution of electrolyte containing a small amount of 
surfactant to improve wetting action. 
 
A thin semi-permeable membrane, stretched over the sensor, isolates the electrodes from the 
environment, while allowing gases to enter. When a polarizing voltage is applied to the sensor 
electrodes oxygen that has passed through the membrane reacts at the cathode causing a current to flow. 
 
The membrane passes oxygen at a rate proportional to the pressure difference across it. Since oxygen is 
rapidly consumed at the cathode, it can be assumed that the oxygen pressure inside the membrane is 
zero. Hence, the force causing the oxygen to diffuse through the membrane is proportional to the partial 
pressure of oxygen outside the membrane. As the oxygen partial pressure varies, so does the oxygen 
diffusion through the membrane. This causes the probe current to change proportionally. 
 
 

PREPARING THE PROBE 

 

MEMBRANE CAP INSTALLATION 

 
The YSI 550A DO probe is shipped with a dry, protective membrane.  Before using the instrument for 
the first time, remove the protective cap and replace it with a new one following these instructions:   
 
1.  Remove the probe sensor guard to access the probe tip. 
2.  Unscrew and remove the old membrane cap and discard. 
3.  Thoroughly rinse the sensor tip with distilled or DI water. 
4.  Fill a new membrane cap with O

2

 probe solution that has been prepared according to the directions 

on the bottle.  Be very careful not to touch the membrane surface.  Lightly tap the side of the 
membrane cap to release bubbles that may be trapped. 

5.  Thread the membrane cap onto the probe.  It is normal for a small amount of electrolyte to 

overflow. 

6.  Replace the probe sensor guard.  
 

MEMBRANE MAINTENANCE 

 
Additional membrane changes will be required over time.  The average replacement interval is 4 to 8 
weeks, although they may last longer if kept clean.  To clean the membrane, use a lint-free cloth, such 
as a Kimwipe, and rubbing alcohol to gently remove the contamination.  In harsh environments, such 
as wastewater, membrane replacements may be required every 2 to 4 weeks. 

Summary of Contents for 550A

Page 1: ...Operations Manual YSI 550A Dissolved Oxygen Instrument...

Page 2: ......

Page 3: ...S OF THE YSI 550A 3 PRINCIPLES OF OPERATION 5 PREPARING THE PROBE 5 DISSOLVED OXYGEN CALIBRATION 6 SALINITY COMPENSATION CALIBRATION 7 PROBE OPERATION 8 PROBE ELECTRODE MAINTENANCE 9 ACCESSORIES REPLA...

Page 4: ......

Page 5: ...gle calibration regardless of which dissolved oxygen display is used and will calibrate in either mode Salinity compensation values can be changed at any time without performing a new calibration A de...

Page 6: ...to the Authorized Service Center selected by YSI Repair or replacement will be made and the product returned transportation prepaid Repaired or replaced products are warranted for the balance of the...

Page 7: ...and salinity calibration and Increases or decreases the value during calibrations and Mode Press at the same time to switch the temperature units between Fahrenheit F and Celsius C and Mode Press at t...

Page 8: ...the instrument by unscrewing the two screws 2 Remove the rubber stopper from the chamber by pulling the tail free of the small hole on the chamber 3 Re attached the rubber stopper to the storage chamb...

Page 9: ...causes the probe current to change proportionally PREPARING THE PROBE MEMBRANE CAP INSTALLATION The YSI 550A DO probe is shipped with a dry protective membrane Before using the instrument for the firs...

Page 10: ...mation can be obtained over the internet or from a local airport or weather station To convert from meters to feet divide by 0 3048 For best results Check calibration with each use and recalibrate as...

Page 11: ...ess and release both the UP ARROW and DOWN ARROW keys at the same time to enter the calibration menu 4 Press the Mode key until mg L is displayed on the right side of the screen for oxygen units Press...

Page 12: ...hrough the sample 3 Allow temperature and dissolved oxygen readings to stabilize 4 Observe Record readings 5 If possible rinse the probe with clean water after each use PRECAUTIONS 1 Membranes last lo...

Page 13: ...ne does not resolve the problem then proceed with the chemical cleaning Mechanical cleaning Sand off the dark layer from the silver anode with 400 grit wet dry sandpaper Wrap the wet sandpaper around...

Page 14: ...the hex screw was removed completely make sure that it is not cross threaded when replaced 7 The YSI 559 DO module is shipped with a dry protective membrane Before using for the first time remove the...

Page 15: ...module Dissolved Oxygen Response Time 95 of end value in 9 seconds Temperature Units Celsius or Fahrenheit user selectable Parameter Compensation Automatic temperature compensation for dissolved oxyge...

Page 16: ...Return system for service 4 Instrument readings are inaccurate A Verify calibration altitude and salinity settings are correct and recalibrate B Probe may not have been in 100 water saturated air duri...

Page 17: ...Main display reads Err or burn and Secondary display reads EEP A Return system for service CONTACT INFORMATION YSI offers a wide range of customer assistance and technical support functions to ensure...

Page 18: ...y from the receiver plug the YSI Instrument into a different outlet so that the computer and receiver are on different branch circuits If necessary the user should consult the dealer or an experienced...

Reviews: