background image

 

YSI  550A

 

Y S I

Environmental

 

Page  9 

 

PROBE ELECTRODE MAINTENANCE  

 

SILVER ANODE 

 
After extended use, a thick layer of AgCl builds up on the silver anode reducing the sensitivity of the 
sensor. The anode must be cleaned to remove this layer and restore proper performance. The cleaning 
can be chemical or mechanical: 
 

• 

Chemical cleaning:

 Remove the membrane cap and rinse the electrodes deionized or 

distilled water.  Soak the entire anode section in a 14% ammonium hydroxide solution for 2 
to 3 minutes (or a 3% solution may be used and soaked overnight for 8-12 hours).  Rinse 
heavily in cool tap water followed by a thorough rinsing with distilled or deionized water. 
The anode should then be thoroughly wiped with a wet paper towel to remove the residual 
layer from the anode. 

 

Warning:  Chemical cleaning should be performed on an as-needed basis, and no more 
often than once a year (or once per six months in wastewater environments).  When 
readings appear unstable or the instrument will not calibrate, first attempt a 
membrane change and recalibrate.  If a new membrane does not resolve the problem, 
then proceed with the chemical cleaning. 

 

• 

Mechanical cleaning:

 Sand off the dark layer from the silver anode with 400 grit wet/dry 

sandpaper. Wrap the wet sandpaper around the anode and twist the probe. Rinse the anode 
with clean water after sanding, and wipe thoroughly with a wet paper towel. 

 

GOLD CATHODE 

 
For correct probe operation, the gold cathode must be textured properly. It can become tarnished or 
plated with silver after extended use. The gold cathode can be cleaned by using the YSI 5238 Probe 
Reconditioning Kit or 400 grit wet/dry sandpaper.  Never use chemicals or abrasives not recommended 
or supplied by YSI.  
 
Using the sanding paper provided in the YSI 5238 Probe Reconditioning Kit, wet sand the gold with a 
twisting motion about 3 times or until all silver deposits are removed and the gold appears to have a 
matte finish. If the cathode remains tarnished, return the probe for service. Wipe the gold cathode 
thoroughly with a wet paper towel before putting on a new membrane cap. 
 

REPLACEMENT OF THE DO ELECTRODES 

 
Should replacement of the DO Electrodes be required, the user may purchase a YSI 559 Replaceable 
DO Module Kit.  The kit includes an instruction sheet, DO sensor module, set screw, and hex key 
wrench. 

 

Summary of Contents for 550A

Page 1: ...Operations Manual YSI 550A Dissolved Oxygen Instrument...

Page 2: ......

Page 3: ...S OF THE YSI 550A 3 PRINCIPLES OF OPERATION 5 PREPARING THE PROBE 5 DISSOLVED OXYGEN CALIBRATION 6 SALINITY COMPENSATION CALIBRATION 7 PROBE OPERATION 8 PROBE ELECTRODE MAINTENANCE 9 ACCESSORIES REPLA...

Page 4: ......

Page 5: ...gle calibration regardless of which dissolved oxygen display is used and will calibrate in either mode Salinity compensation values can be changed at any time without performing a new calibration A de...

Page 6: ...to the Authorized Service Center selected by YSI Repair or replacement will be made and the product returned transportation prepaid Repaired or replaced products are warranted for the balance of the...

Page 7: ...and salinity calibration and Increases or decreases the value during calibrations and Mode Press at the same time to switch the temperature units between Fahrenheit F and Celsius C and Mode Press at t...

Page 8: ...the instrument by unscrewing the two screws 2 Remove the rubber stopper from the chamber by pulling the tail free of the small hole on the chamber 3 Re attached the rubber stopper to the storage chamb...

Page 9: ...causes the probe current to change proportionally PREPARING THE PROBE MEMBRANE CAP INSTALLATION The YSI 550A DO probe is shipped with a dry protective membrane Before using the instrument for the firs...

Page 10: ...mation can be obtained over the internet or from a local airport or weather station To convert from meters to feet divide by 0 3048 For best results Check calibration with each use and recalibrate as...

Page 11: ...ess and release both the UP ARROW and DOWN ARROW keys at the same time to enter the calibration menu 4 Press the Mode key until mg L is displayed on the right side of the screen for oxygen units Press...

Page 12: ...hrough the sample 3 Allow temperature and dissolved oxygen readings to stabilize 4 Observe Record readings 5 If possible rinse the probe with clean water after each use PRECAUTIONS 1 Membranes last lo...

Page 13: ...ne does not resolve the problem then proceed with the chemical cleaning Mechanical cleaning Sand off the dark layer from the silver anode with 400 grit wet dry sandpaper Wrap the wet sandpaper around...

Page 14: ...the hex screw was removed completely make sure that it is not cross threaded when replaced 7 The YSI 559 DO module is shipped with a dry protective membrane Before using for the first time remove the...

Page 15: ...module Dissolved Oxygen Response Time 95 of end value in 9 seconds Temperature Units Celsius or Fahrenheit user selectable Parameter Compensation Automatic temperature compensation for dissolved oxyge...

Page 16: ...Return system for service 4 Instrument readings are inaccurate A Verify calibration altitude and salinity settings are correct and recalibrate B Probe may not have been in 100 water saturated air duri...

Page 17: ...Main display reads Err or burn and Secondary display reads EEP A Return system for service CONTACT INFORMATION YSI offers a wide range of customer assistance and technical support functions to ensure...

Page 18: ...y from the receiver plug the YSI Instrument into a different outlet so that the computer and receiver are on different branch circuits If necessary the user should consult the dealer or an experienced...

Reviews: