background image

 

Page  6

 

Y S I

Environmental

 

YSI  550A 

 

DISSOLVED OXYGEN CALIBRATION 

 
Dissolved oxygen calibration must be done in an environment with known oxygen content.  The YSI 
550A DO Instrument can be calibrated in either mg/L or % saturation.  Sections below include 
instructions on how to calibrate in either mode.   
 

BEFORE YOU CALIBRATE 

To accurately calibrate the YSI 550A, you will need to know the following information: 
 

• 

The approximate salinity of the water you will be analyzing. Fresh water has a salinity of 
approximately zero. Seawater has a salinity of approximately 35 parts per thousand (ppt). If you 
are uncertain what the salinity of the sample water is, use a YSI 30 Salinity-Conductivity-
Temperature instrument to determine a salinity value. 

 

• 

For calibration in % saturation mode, the approximate altitude (in feet) of the region where you are 
located is required.  This information can be obtained over the internet or from a local airport or 
weather station.  To convert from meters to feet, divide by 0.3048. 

 

For best results: 
 

• 

Check calibration with each use and recalibrate as necessary to prevent drift.  Dissolved oxygen 
readings are only as good as the calibration. 

 

• 

Calibrate at a temperature within 

±

10°C of the sample temperature.

 

 

CALIBRATION IN % SATURATION 

 

1. 

Ensure that the sponge inside the instrument's calibration chamber is moist. Insert the probe into 
the calibration chamber. 

 
2. 

Turn the instrument on.  Allow it to warm up and readings to stabilize for about 15-20 minutes. 

 
3. 

Press and release both the 

UP ARROW

 and 

DOWN ARROW

 keys at the same time to enter the 

calibration menu. 

 
4. Press 

the 

Mode

 key until “%” is displayed on the right side of the screen for oxygen units.  Press 

ENTER

 
5. 

The LCD will prompt you to enter the local altitude in hundreds of feet. Use the arrow keys to 
increase or decrease the altitude.  When the proper altitude appears on the LCD, press the 

ENTER

 key. 

  EXAMPLE:

  Entering the number 12 here indicates 1200 feet. 

Summary of Contents for 550A

Page 1: ...Operations Manual YSI 550A Dissolved Oxygen Instrument...

Page 2: ......

Page 3: ...S OF THE YSI 550A 3 PRINCIPLES OF OPERATION 5 PREPARING THE PROBE 5 DISSOLVED OXYGEN CALIBRATION 6 SALINITY COMPENSATION CALIBRATION 7 PROBE OPERATION 8 PROBE ELECTRODE MAINTENANCE 9 ACCESSORIES REPLA...

Page 4: ......

Page 5: ...gle calibration regardless of which dissolved oxygen display is used and will calibrate in either mode Salinity compensation values can be changed at any time without performing a new calibration A de...

Page 6: ...to the Authorized Service Center selected by YSI Repair or replacement will be made and the product returned transportation prepaid Repaired or replaced products are warranted for the balance of the...

Page 7: ...and salinity calibration and Increases or decreases the value during calibrations and Mode Press at the same time to switch the temperature units between Fahrenheit F and Celsius C and Mode Press at t...

Page 8: ...the instrument by unscrewing the two screws 2 Remove the rubber stopper from the chamber by pulling the tail free of the small hole on the chamber 3 Re attached the rubber stopper to the storage chamb...

Page 9: ...causes the probe current to change proportionally PREPARING THE PROBE MEMBRANE CAP INSTALLATION The YSI 550A DO probe is shipped with a dry protective membrane Before using the instrument for the firs...

Page 10: ...mation can be obtained over the internet or from a local airport or weather station To convert from meters to feet divide by 0 3048 For best results Check calibration with each use and recalibrate as...

Page 11: ...ess and release both the UP ARROW and DOWN ARROW keys at the same time to enter the calibration menu 4 Press the Mode key until mg L is displayed on the right side of the screen for oxygen units Press...

Page 12: ...hrough the sample 3 Allow temperature and dissolved oxygen readings to stabilize 4 Observe Record readings 5 If possible rinse the probe with clean water after each use PRECAUTIONS 1 Membranes last lo...

Page 13: ...ne does not resolve the problem then proceed with the chemical cleaning Mechanical cleaning Sand off the dark layer from the silver anode with 400 grit wet dry sandpaper Wrap the wet sandpaper around...

Page 14: ...the hex screw was removed completely make sure that it is not cross threaded when replaced 7 The YSI 559 DO module is shipped with a dry protective membrane Before using for the first time remove the...

Page 15: ...module Dissolved Oxygen Response Time 95 of end value in 9 seconds Temperature Units Celsius or Fahrenheit user selectable Parameter Compensation Automatic temperature compensation for dissolved oxyge...

Page 16: ...Return system for service 4 Instrument readings are inaccurate A Verify calibration altitude and salinity settings are correct and recalibrate B Probe may not have been in 100 water saturated air duri...

Page 17: ...Main display reads Err or burn and Secondary display reads EEP A Return system for service CONTACT INFORMATION YSI offers a wide range of customer assistance and technical support functions to ensure...

Page 18: ...y from the receiver plug the YSI Instrument into a different outlet so that the computer and receiver are on different branch circuits If necessary the user should consult the dealer or an experienced...

Reviews: