26
SYMPTOM
________________________________________________________________
Fluid leaking; appears to be coming from the Monitor Pump.
POSSIBLE CAUSE: Either the coupling between the sample line and sample pump tubing is leaking or a
pump tubing has failed.
CORRECTIVE ACTION: Locate the leak. If inside the pump, refer to Section 8.1 to replace pump
tubing.
SYMPTOM
________________________________________________________________
Sipper aspirates antiseptic instead of sample (sample results read approximately zero).
POSSIBLE CAUSE: Tubing routed incorrectly in solenoid valve.
CORRECTIVE ACTION: Check tubing routing per Section 8.1.2.
NOTE: Replace membrane if damaged by antiseptic.
FOR PROBLEMS NOT IDENTIFIED IN THE ABOVE CHART CONTACT YSI TECHNICAL
SUPPORT FOR ASSISTANCE. REFER TO APPENDIX B FOR PHONE, FAX AND ADDRESS
INFORMATION.