27
CLEANING THE UV SENSOR AND CALIBRATION
Clean the UV sensor cover lens
For precision UV measurement, gentle clean the UV sensor cover lens by the pure water
regularly .
UV Sensor Degradation
Over time, the UV sensor will naturally degrade . The UV sensor can be calibrated with a utility
grade UV meter .
CONSOLE FIRMWARE UPDATE
1 . Unzip the “
Update.zip
” file
2 . Copy the unzipped “
Update
” folder to the
USB mass storage driver .
Important note:
-The USB mass storage driver should be
FAT32
format
-The “
Update
” folder must be the USB mass
storage driver’s root directory)
3 .
Disconnect the main console’s DC power from the power outlet and remove the backup
batteries . Plug the USB mass storage driver to the USB socket at the right side of the main
console .
4 .
Reconnect the main console’s DC power to start the update process
Important note:
-In the meantime, do not remove the USB mass storage driver .
-
If the firmware update still not restart, you can try to press the reset key to trigger the
update process .
5 .
During the update process, the Console status LED will flash in red color and LCD will show
the update status as below:
Updating screen
Update completed and waiting for reboot