![YOKOGAWA YTA610 User Manual Download Page 41](http://html1.mh-extra.com/html/yokogawa/yta610/yta610_user-manual_3854331041.webp)
<4. Diagnostics >
4-9
IM 01C50T01-02EN
Name
bit
Description
NAMUR-107 Mask Available
Initial Setting
Device Specific Status 1
0
Temp NV Warning
M
No mask
1
Int Comm Fail
F
No mask
Device Specific Status 2
0
Sensor1 Failure
F
No mask
1
Sensor2 Failure
F
No mask
2 Sensor1 Short
*1
F
Mask
3 Sensor2 Short
*1
F
Mask
4 Sensor1 Corrosion
*1
M
Mask
5 Sensor2 Corrosion
*1
M
Mask
6
Sensor1 Signal Error
S
No mask
7
Sensor2 Signal Error
S
No mask
Device Specific Status 3
0
Terminal Sensor Fail
F
No mask
1
Backup Sensor1 Fail
M
No mask
2
Backup Sensor2 Fail
M
No mask
3 Sensor Drift
M
Mask
4
Sensor1 Temp Cycle
*1
M
Mask
5
Sensor2 Temp Cycle
*1
M
Mask
Device Specific Status 4
0
Output Too Low
S
No mask
1
Output Too High
S
No mask
2
Sensor1 Temp Low
S
No mask
3
Sensor1 Temp High
S
No mask
4
Sensor2 Temp Low
S
No mask
5
Sensor2 Temp High
S
No mask
6
Amb Temp Low
S
No mask
7
Amb Temp High
S
No mask
Device Specific Status 5
1
LRV Too Low
C
No mask
2
LRV Too High
C
No mask
3
URV Too Low
C
No mask
4
URV Too High
C
No mask
5
Span Too Small
C
No mask
Device Specific Status 14
0
Illegal PV Cfg
C
No mask
1
Illegal Sensor1 Cfg
C
No mask
2
Illegal Sensor2 Cfg
C
No mask
Device Specific Status 15
0
Output Manual Mode
C
No mask
*1:
Applicable only for YTA710.