6-4
IM 01B08D02-02EN
Operating Conditions
Normal operating conditions, transport and storage conditions
Item
Normal Operation
Transport and Storage
Ambient
temperature
0 to 50°C
−20 to 60°C
Ambient humidity
5 to 90% RH
(No condensation)
5 to 95% RH
(No condensation)
Power Supply
voltage (AC) (*1)
80 to 138 V AC (100 V AC and 24 V AC
common power supply)
138 to 264 V AC (220 V AC power supply)
–
Power supply
frequency (AC)
50/60 Hz ±3Hz
–
Power Supply
voltage (DC) (*1)
20 to 132 V DC (100 V AC and 24 V DC
common power supply)
120 to 340 V DC (220 V AC power supply)
–
Continuous
vibration
5 to 14 Hz, amplitude : 0.625 mm or less
14 to 150 Hz, 4.9 m/s
2
or less, orthogonal three-directional, with 2 hours each
Short-time
vibration
14.7 m/s
2
, 15 s or less
Shock
49 m/s
2
(5 G) or less
11 ms or less
Package drop
Within 1 m
Magnetic field
400 A/m or less
Hazardous gas
There shall be no corrosive gas in the location.
Installation
altitude
2000 m above the sea or less
Atmospheric
pressure
86 k to 106 kPa
*1: Safety standards, FM and CSA non-incendive standard compliant conditions apply at the
following rated power supply.
Rated power supply: use of both AC and DC
100 V AC and 24 V DC common power supply
DC drive: 24 - 120 V DC (±10%), no polarity, 750 mA MAX.
AC drive: 100 - 120 V AC (±10%), 50/60 Hz (±3 Hz), 30 VA MAX.
220 V AC power supply
DC drive: 135 - 190 V DC (±10%), no polarity, 110 mA MAX.
AC drive: 220 - 240 V AC (±10%), 50/60 Hz (±3 Hz), 30 VA MAX.
6.1 General Specifications