
Model YPP6200 Analog Input Rate/Totalizer
Instruction Manual
73
Noise Filter (
filter
)
The noise filter is available for unusually noisy signals that cause an
unstable process variable display. The noise filter averages the input
signal over a certain period. The filter level determines the length of time
over which the signal is averaged. The filter level can be set between 2
and 199. The higher the filter level, the longer the averaging time and so
the longer it takes the display to settle to its final value. Setting the filter
level to zero disables the filter function.
Noise Filter Bypass (
bypass
)
The noise filter bypass changes the behavior of the meter so that small
variations in the signal are filtered out but large abrupt changes in the
input signal are displayed immediately. The bypass value determines the
minimum amount of signal change to be displayed immediately. All
signal changes smaller than the bypass value are filtered or averaged by
the meter. The noise filter bypass may be set between 0.1 and 99.9% of
full scale.
Rounding Feature (
round
)
The rounding feature is used to give the user a steadier display with
fluctuating signals. Rounding is used in addition to the filter function.
Rounding causes the display to round to the nearest value according the
rounding selected. See examples below:
Rounding
Selection
Actual
Value
Display
Value
Actual
Value
Display
Value
1 12.022
12.022
12.023
12.023
5 12.022
12.020
12.023
12.025
10 12.024
12.020
12.025
12.030