
App-4
IM WT18O1E-02EN
Code
Message
Chapter or Section
850
Cannot be set or executed at current store state.
To set or execute, reset store.
16.4
852
Stored file is illegal.
Initialize memory before storing.
16.4
854
Waveform display data not found.
—
855
Data destination memory is full.
Saving has been stopped.
—
856
An error has occured while storing.
Storing has been stopped.
—
857
Cannot be set while Master/Slave Synchronized
Measurement is set to Slave.
7.6
858
Store process is in progress now.
Execute or set setting again.
—
859
Cannot convert selected file.
Select a file with an extension of WTS or HDS.
16.3
862
Numeric data not found.
—
863
Cannot be set or executed when different types of elements are installed.
—
864
This wiring cannot be set as the first selected element.
1.1
865
Cannot be set while integration is running.
Stop or reset Integration.
8.3
866
Cannot be set or executed while Auto Print is operating.
Turn off Auto Print from the [PRINT MENU] (SHIFT+PRINT) menu.
19.2
867
Auto Print is not in operation.
Start Auto Print from the [PRINT MENU] (SHIFT+PRINT).
19.2
868
Print out destination must be set to Built-in Printer in order to start Auto Print.
Set [Print to] to Built-in from the [PRINT MENU] (SHIFT+PRINT) menu.
19.2
869
Auto Print funciton is not supported in the current measurement mode or settings.
19.2
870
Auto Print [Interval] setting is invalid.
Set [Interval] time to an appropriate amount from the [PRINT MENU](SHIFT+PRINT)
->[Auto Print Settings] menu.
19.2
871
Attempted made to start Auto Print when the end time had already passed.
Set [End Time] to a future date & time from the [PRINT MENU](SHIFT+PRINT)
->[Auto Print Settings] menu.
19.2
872
Auto print’s print-out has been canceled.
The printer or file system is in action.
19.2
874
Sync source, PLL source or trigger source cannot be set to Ext Clk, while Master/Slave
Synchronization Measurement is set to Slave.
7.6
875
Master/Slave Synchronization Measurement cannot be set to Slave, while sync
source, PLL source or trigger source is set to Ext Clk.
7.6
876
Can not calculate from present point value.
3.1 or 4.1
877
Can not set 0 to count.
15.1 or 16.1
879
Can not set or execute while recording high speed data.
Stop measurement and wait for file status "Ready".
15.4
880
Cannot be set or executed while initialization.
Wait until status changes to "Ready".
15.4
881
Cannot be set or executed while measurement is in progress.
To set or execute, "Stop" measurement.
15.4
882
Stopped measurement. Detection error of measuring interval signal.
Check External Sync (MEAS START) input.
4.4*
883
Cannot be set or executed in High Speed Data Capturing Mode.
—
884
Can not set wiring to 1P3W/3P3W in High Speed Data Capturing Mode.
Select a different wiring.
1.1
885
Cannot be set or executed in High Speed Data Capturing Mode.
Set or execute in Normal Measurement Mode.
Appendix 10*
886
Cannot be set or executed to same current ranges, due to different types of elements
are installed or external current sensor settings are not same.
1.3
887
Cannot start integration. Turn off Independent Element setting by the [WIRING] menu,
or switch the measurement ranges to fixed ranges.
1.1
888
Cannot start the integration. Turn off Independent control by the [INTEG] menu or turn
off Auto of the Data Update Inerval by the [UPDATE RATE] menu.
8.1 or 1.15
889
Setting and execution is not available when auto-ranging is set to ON.
1.2
* Getting started guide, IM WT1801E-03EN
Appendix 1 Messages and Corrective Actions