5-31
IM 760101-11E
Commands
5
:DISPlay:WAVE:{U<x>|I<x>|SPEed|TORQue}
Function
Turns ON/OFF the {voltage|current|rotating
speed|torque} waveform or queries the current
setting.
Syntax
:DISPlay:WAVE:{U<x>|I<x>|SPEed|
TORQue} {<Boolean>}
:DISPlay:WAVE:{U<x>|I<x>|SPEed|
TORQue}?
<x> = 1 to 6
Example
:DISPLAY:WAVE:U1 ON
:DISPLAY:WAVE:U1? -> :DISPLAY:WAVE:
U1 1
Description
{SPEed|TORQue}
is valid only when the motor
evaluation function (/MTR option) is installed.
* Function Selection (<Function>) List
(1) Functions in the Normal Measurement Mode
Applicable commands
:AOUTput[:NORMal]:CHANnel<x>
:DISPlay[:NUMeric]:NORMal:FCURsor
:DISPlay[:NUMeric]:NORMal:ITEM<x>
:DISPlay:TRENd:NORMal:ITEM<x>[:FUNCtion]
:FILE:SAVE:NUMeric:NORMal:...
:HCOPy:PRINter:DLISt:NORMal:...
:NUMeric[:NORMal]:ITEM<x>
:STORe:NUMeric:NORMal:...
Function names used: Function names used on the
in communications
menu
(Numerical display header name)
URMS
: Urms
UMN
: Umean
UDC
: Udc
UAC
: Uac
IRMS
: Irms
IMN
: Imean
IDC
: Idc
IAC
: Iac
P
: P
S
: S
Q
: Q
LAMBda
:
λ
PHI
:
φ
FU
: FreqU
(fU)
FI
: FreqI
(fI)
UPPeak
: U+peak
(U+pk)
UMPeak
: U-peak
(U-pk)
IPPeak
: I+peak
(I+pk)
IMPeak
: I-peak
(I-pk)
CFU
: CfU
CFI
: CfI
FFU
: FfU
FFI
: FfI
Z
: Z
RS
: Rs
XS
: Xs
RP
: Rp
XP
: Xp
PC
: Pc
TIME
: I-Time
WH
: Wp
WHP
: Wp+
WHM
: Wp-
AH
: q
AHP
: q+
AHM
: q-
ETA
:
η
SETA
: 1/
η
F1
: F1
F2
: F2
F3
: F3
F4
: F4
DURMS
:
∆
Urms
DUMN
:
∆
Umean
DUDC
:
∆
Udc
DUAC
:
∆
Uac
DIRMS
:
∆
Irms
DIMN
:
∆
Imean
DIDC
:
∆
Idc
DIAC
:
∆
Iac
SPEed
: Speed
TORQue
: Torque
SYNC
: SyncSpd
SLIP
: Slip
PM
: Pm
MAETa
:
η
mA
MBETa
:
η
mB
Note
The measurement functions below are not related to the primary
element, but you must use a setting of 1 if elements must be set
up useing the D/A output command (AOUTput:NORMal), the
numerical display command for normal measurement
(DISplay:NUMeric:NORMal), the trend display command for
normal measurement (DISplay:TRENd:NORMal), or the
numerical data output command for normal measurement
(NORMal).
The functions are: ETA (efficiency), SETA (1/efficiency), SPEed
(motor rotation speed), TORQue (moror torque), SYNC (motor
synchronous speed), SLIP (motor slip), PM (motor output),
MAETa (total efficiency).
(2) Functions in the Harmonic Measurement Mode
(Numerical Display)
Applicable commands
:AOUTput:HARMonics:CHANnel<x>
:DISPlay[:NUMeric]:HARMonics:ITEM<x>
:DISPlay:TRENd:HARMonics:ITEM<x>
[:FUNCtion]
:NUMeric:HARMonics:ITEM<x>
Selections used in : Function names used on the
communications
menu
(Numerical display header name)
U
: U
I
: I
P
: P
S
: S
Q
: Q
5.5 DISPlay Group
Summary of Contents for wt1600
Page 1: ...Digital Power Meter Communication Interface IM 760101 11E 4th Edition ...
Page 131: ...7 2 IM 760101 11E 7 2 Sample Program Image ...
Page 138: ...Sample Program 7 9 IM 760101 11E 7 7 4 Output of Normal Measurement Data ...
Page 142: ...Sample Program 7 13 IM 760101 11E 7 7 5 Output of Harmonic Measurement Data ...