<4. PARAMETER SETTING>
4-7
IM 12E01A06-02E
(2)
HOLD Mode: {HOLD}
In this mode whether the analog output is held or not in measurement mode, can be selected. This
function is available when the hold function is set to “1: Enabled” in SET HOLD mode at the setting
level. If the function is set to “0: Disabled,” this mode will be skipped.
F4-5E.ai
[YES]
[YES]
[NO]
Hold enabled in
measurement mode
Hold disabled in
measurement mode
HOLD
HOLD
Measurement
mode
HOLD status indicator turns on.
(Flashing)
Figure 4.3
HOLD Mode Flow Chart
If the hold function has been set to “1: Enabled” in SET HOLD mode at the setting level, the instrument
automatically goes into this HOLD mode when it returns from any level to measurement mode, with
the exception of DISPLAY mode.
(3)
RANGE Mode: {*RANGE}
In this mode, an output range corresponding to analog output 1 and 2 can be set. Enter turbidity
values corresponding to a 4 mA (or 0 mA) output signal (zero point) and a 20 mA output signal (span
point).
[YES]
F4-6E.ai
[NO]
[YES]
[YES]
[YES]
[YES]
[ENT]
[YES]
[ENT]
[NO]
[NO]
[YES]
[ENT]
[
]
[
]
[
]
[
]
[
]
[
]
[
]
[
]
[
]
[
]
*RANGE
*LOCAL
*FIXR.2
*FIXR.1
XXXX.XX
*ZERO
XXXX.XX
*SPAN
XXXX.XX
*SPAN
XXXX.XX
*ZERO
X
*LCL.1
X
*LCL.2
*1: Skipped if analog output range switching selection {*RNGPR} is set to “1: Analog output 1”
in CODE 30 at service level.
*2: Skipped if analog output range switching selection {*RNGPR} is set to “2: Analog output 2”
in CODE 30 at service level.
*3: Displayed only if analog output range switching selection {*RNGPR} is set to “1: Analog
output 1” or “2: Analog output 2” and range switching function selection {*RSET} is set to
“0: Manual” in CODE 30 at service level.
*4: Displayed only if analog output range switching selection {*RNGPR} is set to “1: Analog
output 1” in CODE 30 at service level.
*5: Displayed only if analog output range switching selection {*RNGPR} is set to “2: Analog
output 2” in CODE 30 at service level.
*6: Zero and span points should be set together. Only after span point has been entered, both
zero and span points are accepted. Cancellation by pressing [MODE] key during the
procedure will not change either of the points.
*1
*2
*3
*6
*6
*4
*5
Local range for output 1
X: 0(Range A), 1(Range B)
or 2(Range C)
Local range for output 2
X: 0(Range A), 1(Range B)
or 2(Range C)
Figure 4.4
RANGE Mode Flow Chart