<4. PARAMETER SETTING>
4-5
IM 12E01A06-02E
Code
Setting Parameter
Display
Setting Range
Default
CODE37
Time constant setting
Time constant in measurement
*TC
000 to 120[s]
020
(s)
Time constant in maintenance
*TC-M
000 to 120[s]
006
(s)
CODE40 Contact output S1 function selection
Function of contact output S1
* S1
0:None , 1: Alarm,
2: Automatic calibration/
washing, 3: Maintenance
1
: Alarm
CODE41 Contact output S2 function selection
Function of contact output S2
* S2
0:None , 1: Alarm,
2: Automatic calibration/
washing, 3: Maintenance
3
: Maintenance
CODE44 Delay time/hysteresis setting
Delay time
* D.TIME
000 to 199[s]
000
(s)
Hysteresis
* HYST
000 to 100[%]
002
(%)
CODE50
Auto return function selection
Auto return function enabled or disabled
*RET
0: Disabled, 1: Enabled
0
: Disabled
CODE51
Alarm setting function selection
Alarm setting function enabled or disabled
* MODE
0: Disabled, 1: Enabled
0
: Disabled
CODE52 Password setting
Password
*PASS
0:No password, 1:111, 2:333,
3:777,4:888, 5:123, 6:957,
7:331, 8:546, 9:847
0.0.0
CODE54 Negative nondisplay/nonoutput selection
Negative nondisplay/nonoutput enabled or
disabled
*MINUS
0: Disabled, 1: Enabled
0.0
: Disabled.
Disabled
CODE61
Turbidity unit selection
Turbidity unit
*UNIT
0: NTU, 1: FNU, 2: mg/l
0
: NTU
CODE64 Software version display
Software version display
*VER
1.00
CODE66 Error “E2xx” error level selection
Error levels of errors “E201” to “E206”
*201-6
0: Disable, 1: Level 1, 2: Level
2
1.1.2.2.2.2
CODE67 Error “E3xx” error detection selection
Error detection for “E301” to “E306”
*301-6
0: Disabled, 1: Enabled
1.1.1.1.1.1
Error detection for “E307”
*307
0: Disabled, 1: Enabled
1
Error detection for “E311” to “E316”
*311-6
0: Disabled, 1: Enabled
1.1.1.1.1.1
Error detection for “E317”
*317
0: Disabled, 1: Enabled
1
CODE71
Analog output test
Start test for output 1
*AO1.T
Test current flowing
End
*END
Start test for output 2
*AO2.T
Test current flowing
End
*END
CODE72 Contact output test
Test 1
*DO.T1
0: Non-action, 1: Action
Last measured
value held
Test 2
*DO.T2
0: Non-action, 1: Action
Last measured
value held
CODE73 Contact input test
Test
*DI.T
CODE79 Initialization
Start
*LOAD
Default loading
*WAIT
Table 4.3
Setting Parameters at Service Level (2/2)