<4. PARAMETER SETTING>
4-3
IM 12E01A06-02E
4.3 Setting Level
There are modes where parameters regarding output signals and contact outputs can be set.
F4-3E.ai
[YES]
[ * ] For at least second
[NO]
[YES]
[YES]
[NO]
[YES]
[NO]
[NO]
HOLD
*SETP
SETPOINTS Mode
*RANGE
RANGE Mode
*HOLD
SET HOLD Mode
*SERV
Service Level
Measurement
mode
*1: Skipped unless either of functions of contact output S1 or S2 is set to “1:
Alarm” in Code 40 or 41 at service level.
*2: Skipped if hold function is set to “*H.OFF: Disabled” in SET HOLD mode at
setting level.
*3: For service level, see Section 4.4.
*1
*3
*2
Table 4.2
Setting Parameters at Setting Level
Mode
Setting Parameter
Display
Setting Range
Default
SETPOINTS
Alarm setting
* SETP
*1
High alarm setpoint
* S.TB-H
–010.0 to 2200.0[NTU]
*1
2200.0
(NTU)
Low alarm setpoint
* S.TB-L
–010.0 to 2200.0[NTU]
*1
-010.0
(NTU)
RANGE
Analog output range
setting
* RANGE
Fixed range for output 1,
or fixed range for output 2
* FIXR.1 or
*FIXR.2
Zero point
* ZERO
0000.00 to 2000.00[NTU]
0000.00
(NTU)
Span point
* SPAN
0000.00 to 2000.00[NTU]
0100.00
(NTU)
Local range selection
* LOCAL
Local range for output 1,
or local range for output 2
*LCL.1 or
*LCL.2
0: Range A, 1: Range B, 2: Range C
0
:Range A
SET HOLD
Hold parameter setting
* HOLD
Hold function enabled, or
hold function disabled
* H.ON or
*H.OFF
*H.ON
(Enabled)
*H.OFF
(Disabled)
* H.ON
:Enabled
Last measured value, or
fixed value
* H.LST or
* H.FIX
*H.LST
(Last measured value)
*H.FIX
(Fixed value)
* H.LST
:Last
measured value
Fixed value for output 1
* H.mA1
02.0 to 22.0[mA]
22.0
(mA)
Fixed value for output 2
* H.mA2
00.0 to 22.0[mA](0-20mA)
02.0 to 22.0[mA](4-20mA)
22.0
(mA)
CAL/WASH
Not available
SERVICE
*2
*1:
Skipped if alarm setting function is set to “0: Disabled” in Code 51 at service level, or if either function of contact output S1 or S2
is not set to “1: alarm” in Code 40 or 41 at service level.
*2:
For service level, see Section 4.4.