<4. PARAMETER SETTING>
4-4
IM 12E01A06-02E
4.4 Service Level
The TB750G turbidimeter has various functions. At this level there are modes where functions can be
selected according to need for the operation.
Table 4.3
Setting Parameters at Service Level (1/2)
Code
Setting Parameter
Display
Setting Range
Default
CODE02 Input voltage display
Transmitted light detection element input display
* IN1
Scattered light detection element input display
* IN2
End
* END
CODE08 Spike detection parameter setting
Spike detection function enabled or disabled
* SPIKE
0: Disabled, 1: Enabled
0
:Disabled
Limit value
* LIMIT
000.000 to 999.999
999.999
(NTU)
Hold time
* HLD-T
005 to 600[s]
030
(s)
Sampling time
* SMP-T
001 to 600[s]
030
(s)
CODE11 Zero calibration factor display
Zero calibration factor display
*CAL.A
Depends on MS
Code
CODE12
Slope display
Slope display
* SL
100.0
(%)
CODE13 Zero correction factor setting
Zero correction factor setting
*CAL.B
–9.000 to 09.000[NTU]
00.000
(NTU)
CODE14 Sensitivity correction factor setting
Sensitivity correction factor setting
*CAL.K
0.2500 to 4.0000
1.0000
CODE16 Standard solution/check block selection
Standard solution or check block
*CAL.TP
0: Standard Solution, 1: Check
Block
1
:Check Block
CODE17 Check block turbidity setting
Check block turbidity
*PLATE
0.001 to 2000[NTU]
CODE30
Analog output range
switching parameter setting
Range switching output selection
* RNGPR
0: Fixed, 1: AO1, 2: AO2
0
:Fixed
Range switching function selection
*RSET
0: Local, 1: Automatic,
2: Remote
2RNG, 3: Remote 3RNG
1
:Automatic
Range A setting
* RSET.A
Zero point for range A
* ZERO
0000.00 to 2000.00[NTU]
0000.00
(NTU)
Span point for range A
*SPAN
0000.00 to 2000.00[NTU]
0010.00
(NTU)
Range B setting
* RSET.B
Zero point for range B
*ZERO
0000.00 to 2000.00[NTU]
0000.00
(NTU)
Span point for range B
* SPAN
0000.00 to 2000.00[NTU]
0100.00
(NTU)
Range C setting
* RSET.C
Zero point for range C
* ZERO
0000.00 to 2000.00[NTU]
0000.00
(NTU)
Span point for range C
* SPAN
0000.00 to 2000.00[NTU]
1000.00
(NTU)
Automatic range switching point
*AUTOR
070 to 100[%]
080
(%)
CODE33
Analog output 2 range selection
Range selection for output 2
*mA2
0: 4.0-20.0mA, 1: 0.0-20.0mA
0
:4.0-20.0mA
Fixed value for hold
*H.mA2
00.0 to 22.0[mA](0-20mA)
02.0 to 22.0[mA](4-20mA)
22.0
(mA)
Fixed value for hold during failure
*FH.mA2
00.0 to 22.0[mA](0-20mA)
02.0 to 22.0[mA](4-20mA)
22.0
(mA)
CODE35 Hold during failure parameter setting
Hold function enabled or disabled
* FHOLD
0: Disabled, 1: Enabled
1
: Enabled
Last measured value, or fixed value
*FH.LST /
*FH.FIX
*FH.LST
(Last measured
value)
*FH.FIX
(Fixed value)
*FH.FIX:
Fixed
value
Fixed value for output 1
Fixed value for output 2
*FH.mA1
*FH.mA2
02.0 to 22.0[mA]
00.0 to 22.0[mA](0-20mA)
22.0
(mA)
22.0
(mA)
02.0 to 22.0[mA](4-20mA)