3-47
TI04L55B01-01EN
3.3.15
Multi-batch Function (/BT option)
The Multi Batch function allows you to record and save the measured data from multiple batches
separately (per batch). You can also apply the Multi Batch function to multiple consecutive
processes.
Features
Start and stop recording separately for each batch and create data files for each batch.
Equipment 1
Equipment 2
Equipment 12
Equipment 6
Data file
Record start
Record stop
Batch group 12
Batch group 1
Batch group 2
Batch group 6
GM
Record start
Recor stop
Record start
Record start
Record stop
Record stop
• • •
• • •
• • •
• • •
Data file
Data file
Data file
l
Main Specifications
Item
Specification
Number of multi batches
GM10-1: 6 max., GM10-2: 12 max.
Batch single operation
Memory start/stop, math
reset, message writing
Batch overview operation
Computation start/stop, report start/stop, manual sampling, setup data
save/load
Scan interval
500ms, 1 s, 2 s, 5 s (common to all batches)
Data type
Display or event only. Trigger mode not available for event data.
Recording interval
Common to all batches
Data file
Display or event data file created for each batch.
Number of display groups
GM10-1: 6 max. per batch (Number of channels per group: 20)
GM10-2: 12 max. per batch (Number of channels per group:20)
Batch single settings
Group, trip line, file header, data file name, text field, batch number, lot
number
• For details, see the Multi-Batch Function (/BT) User’s Manual.