![YOKOGAWA SC202G(S) User Manual Download Page 64](http://html1.mh-extra.com/html/yokogawa/sc202g-s/sc202g-s_user-manual_3789476064.webp)
11-4 Appendix
11-6. User setting table
FUNCTION
SETTING DEFAULTS
USER SETTINGS
Parameter specific functions
01
*SC.RES
0
SC
02
*4-Elec
0
2-Elec.
03
*0.10xC
0.10xC Factor
1.000
/cm
04
*AIR
05
*POL.C.K
1
On
Temperature measuring functions
10
*T.SENS
0
Pt1000
11
*T.UNIT
0
°C
12
*T.ADJ
None
Temperature compensation functions
20
*T.R.°C
25
°C
21
*T.C.1
2.1
%/°C
*T.C.2
2.1
%/°C
22
*MATRX
None, see 5-2-5
23
*T1°C
T. range
See sep. table, 11-2
24
*L1xT1
Cond. C1
See sep. table, 11-2
25
*L2xT1
Cond. C2
See sep. table, 11-2
26
*L3xT1
Cond. C3
See sep. table, 11-2
27
*L4xT1
Cond. C4
See sep. table, 11-2
28
*L5xT1
Cond. C5
See sep. table, 11-2
mA outputs
31
*OUTP.F
0
Linear S.C.
32
*BURN
0
No Burn
35
*TABL1
21 pt table
see code 31, 11-1