IM 12D7B3-E-H
5-10 Parameter setting
5-3-2. Temperature measuring functions
Code 10
T.SENS
Selection of the temperature compensation sensor. The default selection is the
Pt1000 Ohm sensor, which gives excellent precision with the two wire connec-
tions used. The other options give the flexibility to use a very wide range of other
conductivity/resistivity sensors.
Code 11
T.UNIT
Celsius or Fahrenheit temperature scales can be selected to suit user preference.
Code 12
T.ADJ
With the process temperature sensor at a stable known temperature, the
temperature reading is adjusted in the main display to correspond. The calibra-
tion is a zero adjustment to allow for the cable resistance, which will obviously
vary with length. The normal method is to immerse the sensor in a vessel with
water in it, measure the temperature with an accurate thermometer, and adjust
the reading for agreement.