IM 12D7B3-E-H
IM 12D7B3-E-H
5-2 Parameter setting
5-2. Commissioning mode
5-2-1. Introduction
In order to obtain peak performance from the EXA SC202, you must set it up for each custom application.
Output ranges
: mA output is set as default to 0-1 mS/cm or 0-19.99 M
Ω
.cm.
For enhanced resolution in more stable measuring processes, it may be desirable to
select for example 5-10 µS/cm range.
Hold
: The EXA SC202 transmitter has the ability to “HOLD” the output during maintenance peri-
ods. This parameter should be set up to hold the last measured value, or a fixed value to
suit the process.
Temp1/2
: First and second temperature compensation types and values. (see also section 5-2-4)
* NaCl is the default compensation and is used for neutral salt solutions. Strong solutions
of salts are compensated, as are process waters and pure, and ultrapure water.
* TC temperature coefficient compensation uses a linear temperature compensation fac-
tor. This can be set by calibration or configuration.
* Matrix compensation is an extremely effective way of compensation. Choose from
standard matrix tables, or configure your own to exactly suit your process.
Service
: This selection provides access to the service menu.
What follows are pictorial descriptions of typical frontplate pushbutton sequences for each parameter set-
ting function. By following the simple YES/NO prompts and arrow keys, users can navigate through the
process of setting range, hold and service functions.
Summary of Contents for SC202G(S)
Page 2: ......
Page 24: ...IM 12D7B3 E H ...
Page 48: ...IM 12D7B3 E H ...
Page 52: ...IM 12D7B3 E H ...
Page 54: ...IM 12D7B3 E H ...
Page 60: ...IM 12D7B3 E H ...