IM 12D7B3-E-H
2-5 Specifications
Unclassified Location
Output
Supply
+
_
G
Classified Location
Protective
earth
+
_
FM Approved
Power supply
(HART compatible)
Intrinsically safe design
FM Class I. Div. 1, Group ABCD: T4 for ambient temperature < 55 ºC
T6 for ambient temperature < 40 ºC
EXA SC202S analyser
see text below
For electrical data:
Classified Location
Unclassified Location
+
_
G
Protective
earth
Protective
earth
+
_
Load
Resistance
24 volts DC Nominal
Supply Voltage.
Safety barrier or power supply
(HART compatible)
Intrinsically
safe
design
FM
Approved
FM Class I. Div. 1, Group ABCD: T4 for ambient temperature < 55 ºC
T6 for ambient temperature < 40 ºC
EXA SC202S analyser
with R
int
= 300 Ω
For electrical data:
see text below
SENSOR(S)
terminal 11-16
Max. cable length: 60 mtr.
Cable diam.: 3 to 12mm
SENSOR(S)
terminal 11-16
Max. cable length: 60 mtr.
Cable diam.: 3 to 12mm
Figure 1
Figure 2
Suitable values are:
Voc or Vt ≤ 31.5 Volt DC
lsc or lt ≤ 100 mA
Psc or Pt ≤ 1.2 Watt
Suitable values are:
Voc or Vt ≤ 31.5 Volt DC
Isc or It ≤ 100 mA
•
Electrical data of the EXA SC202S:
- Supply circuit (ter and -):
- Sensor input circuit (terminals 11 through 16):
Maximum input voltage V
max
= 31.5 V.
-
Maximum output voltage V
t
= 14.4 V.
Maximum input current I
max
= 100 mA.
-
Maximum output current I
t
= 12.8 mA.
Maximum input power P
i
= 1.2 W.
-
Maximum allowed external capacitance C
a
= 103 nF.
Effective internal capacitance C
i
= 22 nF.
-
Maximum allowed external inductance L
a
= 200 mH.
Effective internal inductance L
i
= 22 µH.
•
If Hand Held Terminal (HHT) is not connected to the power supply lines of the EXA SC202S (see figure 1):
Any FM Approved barrier or power supply may be used that meets the following requirements.
V
oc
or V
t
≤
31.5 V; I
sc
or I
t
≤
100 mA; C
a
≥
22 nF + C
cable
; L
a
≥
22 µH + L
cable
If HHT is connected to the power supply lines of the EXA SC202S (see figure 2):
The Hand Held Terminal must be FM Approved. Refer to the manufacturers control drawing of the HHT and the barrier/power
supply to determine the cable parameters.
(V
oc
or V
t
) + V
HHT
≤
31.5 V; (I
sc
or I
t
) + I
HHT
≤
100 mA; C
a
≥
22 nF + C
cable
+ C
HHT
; L
a
≥
22 mH + L
cable
+ L
HHT
When installing this equipment, follow the manufacturer’s installation drawing.
Installation should be in accordance with ANSI/ISA RP 12.06.01 "Installation of Intrinsically Safe Systems for Hazardous
(Classified) Locations" and the National Electrical Code (ANSI/NFPA 70).
Control equipment connected to the barrier/power supply must not use or generate more than 250 Vrms or Vdc.
•
Resistance between Intrinsically Safe Ground and earth ground must be less than 1.0
Ω
WARNING
- Substitution of components may impair Intrinsic Safety
To prevent ignition of flammable or combustible atmospheres, disconnect power before servicing or read, understand and
adhere to the manufacturer’s’live maintenance procedures.